Brewer
FLEX
Access
Exam Table
Printed in USA © 2015
12
Document # 2102861 RevA
Figure 21. Foot Controls
OPERATION (continued)
TABLE UP
(Raises the
seat upward)
TABLE DOWN
(Lowers the seat
downward)
MODELS 5700 & 5701
MODELS 5800 & 5801
TABLE UP
(Raises the
seat upward)
TABLE DOWN
(Lowers the
seat downward)
BACK DOWN
(Lowers the back)
BACK UP
(Raises the back)
STOP BUTTON (RED)
(Stops all functions
when pressed)
RETURN TO CHAIR BUTTON
(BLUE)
(Returns table to chair
position when pressed)
GREEN LIGHT
(Lights when power
to the foot control is ON)
STOP
RETURN
TO CHAIR
M
odels
5700 & 5701
Models 5700 and 5701 are equipped with a single pedal foot control that controls the raising and lowering of the seat.
Models
5800 & 5801
Models 5800 and 5801 are equipped with a dual pedal foot control that controls the raising and lowering of the seat,
and the raising and lowering of the seat back. In addition, the foot control features the following:
Return to Chair: When pressed, the table automatically returns to the low chair position. The seat back will rise to the full
up position and the table will lower to the full down position and stop. If the Emergency Stop or either pedal is pressed
while the table is in motion, the system will return to manual mode and all motion will stop.
Stop: When pressed, stops the Return to Chair function.
Green Light: (2 pedal foot controls only) indicates that there is power to the foot control.
Unattended Return to Chair Operation
Remain with and observe patient during Return to Chair operation.
Return to Chair feature may need to be momentarily interrupted
for patient to readjust him or herself, or for hazards to be removed
before resuming motion. Press the Return to Chair button to
resume motion only when patient has situated him or herself, or
hazards have been removed.
NOTICE