Checking the version
: the indicator software version is shown
5.3 Calibration Menu
The scale is configured from the factory to match the specified settings for each unit, as defined by the
product specifications and sales brochure. Modification of the settings can be accomplished by altering
the calibration settings below.
Calibration and/or configuration of calibration settings of your scale should be
accomplished by a trained service technician using certified weights to ensure proper
operation and accuracy. Calibration is not covered under warranty.
5.3.1 Calibration Options
In general weighing mode, press the
TARE
and
ZERO/ON/OFF
keys. Release the keys when the
“CAL” message pops up to open the Calibration Menu. The number of calibrations is shown.
Navigate the menu using the keys, press the
TARE
key to confirm or the
ZERO/ON/OFF
key to cancel the operation / exit the menu.
CAL
SubMenu1
SubMenu2
Option
Description
ZERO
Adjusts the CAL.P0 point only.
LINE
CAL.P0
Stores the Zero calibration point (mandatory).
CAL.P1
Adds the cal. point 1 (mandatory). Use a
standard weight over 10% F.S.
END.Y
YES / NO
Saves and exits the calibration points:
Yes
= go to CAL.END;
No
= add another calibration point
CAL.P2
Adds the cal. point 2 (optional). Use a standard
weight
over 10% F.S. and load used in
CAL.P1
.
END.Y
YES / NO
Saves and exits the calibration points:
Yes
= go to CAL.END;
No
= add another calibration point
CAL.P3
Linear calibration point3: do third weight point
calibration, standard weight must be over
10%FS and be larger than it in
CAL.P2
, this
point can be omitted.
GEO
16
Do not use.