AWT35-100107 Rev AB
22/30
The number of time that the indicator has been configured is displayed.
Navigate the menu using the keys, press the
TARE
key to confirm or the
ZERO/ON/OFF
key to cancel the operation / exit the menu.
CONFIG
SubMenu1 SubMenu2 Option
Default
Description
RESET
YES /
NO
NO
Resets the configuration to factory
setting.
REGULA
NONE
NONE
Do not use.
See SBI 210-LCD Operation Manual.
USA
CANADA
EUROPE
PRIM.N
100
~
100000
2000
(USA)
2500
(UK/EU)
Sets the number of divisions (div) of the
primary unit of measure (lb).
PRIM.D
0.0001
~
50
0.5
(USA)
0.2
(UK/EU)
Sets the weight division (1 div = 0.5 lb)
value of the first unit of measure.
The division value of the secondary unit of
measure (1 div = 0.2 kg) is automatically
determined by the indicator according to
the first unit.
PRIM.Ut
kg / lb
lb
(USA)
kg
(UK/EU)
Selects the primary unit of measure: kg or
lb.
The default calibration unit is the primary
unit.
SECND.N
100
~
125000
2500
(USA)
2000
(UK/EU)
Sets the number of divisions (div) of the
secondary unit of measure (kg).
Max 1.25*(PRIM.N)
MOTION
1
~
255
4
Sets the in-motion weight divisions range:
1
~
255
= ±0.25div * (1~255)
OVER.LD
0
~
100
0
Sets the overload weight limit:
0
= FS + 9 div
1~100 = 101%FS ~ 200%FS
UNITS
kg
YES
/
NO
YES
Sets the active unit of measure:
Yes
= enable the unit;
lb
YES
/
NO
YES