AWT35-100107 Rev AB 

 

13/30 

 

4.3  Comparison Function 

The DS1000 Comparison mode can be used to quickly check the acceptability or unacceptability of an 

item’s weight / count  i.e., check-weighing / check-counting.  This function is available in weighing, 

percent  and counting modes.  Open the User  Submenu 1 

BEEP

  to enable  the audible alarm. The 

COMPAR function can be disabled in the 

Config Menu

. 

 

1.  Power up the scale and zero the display, if necessary, by pressing the 

ZERO/ON/OFF

 

key. Be 

sure the scale is displaying weight in your preferred unit of measure. Press the 

UNIT

  key, if 

necessary. 

 

2.  Press and hold the 

FUNC

 key for three seconds. Toggle the weighing modes by pressing the 

 

 keys and select either the WEIGH, PERCNT or COUNT mode. 

 

3.  Place the item(s) to be weighed/counted on the scale platform 

 

If the weight/pieces is beyond the upper limit, the

 HI 

annunciator lights 

 

If the weight/pieces is within the limits, the

 OK 

annunciator lights 

 

If the weight/pieces is below the lower limit, the

 LO 

annunciator lights 

 

4.  Remove the item(s) from the platform and repeat the steps above for other 

Compar

 weighments. 

In comparison mode, the function of Hold, Print, Acc, Data, Tare and, Zero keys are available. 

 

4.3.1  Setting HI / OK /LO Limits 

1.  In weighing or counting mode, press and hold the 

DATA

 key. Release the key when the “COMP” 

message pops up. Press the 

TARE

 key to confirm. 

 

2.  The 

HI 

annunciator lights. Adjust the upper limit pressing the 

 

 

 keys. Press the 

TARE

 key. 

 

3.  The 

LO 

annunciator lights. Adjust the lower limit pressing the 

 

 

 keys. Press the 

TARE

 key 

to confirm. 

 

4.  The indicator stores the limits and returns to weighing / counting mode screen. 

 

4.4  Counting Mode [Pcs] 

The DS1000 Counting mode has been designed to allow the operator to easily carry out basic sampling 

and counting routines on this scale. The COUNT mode can be disabled in the 

Config Menu

. 

 

1.  Power up the scale and zero the display, if necessary, by pressing the 

ZERO/ON/OFF

 

key. Be 

sure the scale is displaying weight in your preferred unit of measure. Press the 

UNIT

  key, if 

necessary. 

 

2.  Press and hold the 

FUNC

 key for three seconds. Press the 

 

 keys to navigate the options. 

3.  Select the COUNT option and press the 

TARE

 key to confirm. 

The 

Pcs

 annunciator flashes and the last stored piece weight is recalled. 

4.  Place the item(s) to be counted on the scale platform. 
5.  The number of parts is displayed in the Pcs window. 
6.  Remove the item(s) from the platform and repeat the steps above for other 

Count 

weighments. 

In counting mode, the function of Hold, Print, Acc, Data, Tare and, Zero keys are available. 

 

4.4.1  Sampling – Piece Weight 

1.  In counting mode, press the 

DATA

 key. Release the key when the “SPL.---” message pops up. 

2.  Place a known quantity of items to be counted on the scale platform. Press the 

TARE

 key to confirm. 

 

The “INP.PCS” message pops up. 

Summary of Contents for DS 1000-LCD

Page 1: ...Model DS 1000 LCD Operator Manual AWT35 100107 Rev AB...

Page 2: ...roduct names used within this document are trademarks or registered trademarks of their respective holders Groupe de soci t s Avery Weigh Tronix 2021 Tous droits r serv s Aucune partie de cette public...

Page 3: ...ighing 9 3 2 Tare Weighing 9 3 3 Changing Units 9 3 4 Changing Weighing Mode 9 3 5 Print Function Com Ports 9 4 Weighing Modes 11 4 1 Hold Function 11 4 1 1 Automatic Hold Mode Default 11 4 1 2 Averag...

Page 4: ...20 5 3 1 Calibration Options 20 5 3 2 Calibration Procedure 21 5 4 Service Configuration Menus 21 5 4 1 Configuration Options 21 6 Specifications 25 Appendix Print Out Formats FMT 25 Manual revision...

Page 5: ...he mains lead must be connected to a supply outlet with a protective earth contact The electrical supply at the socket outlet must provide over current protection of an appropriate rating For your pro...

Page 6: ...Guidelines To get the best performance from the scale link the platform and indicator connectors and place the DS 1000 LCD scale in a location that will not degrade its accuracy Try to avoid placing...

Page 7: ...Tare is 0 NET Indicates a Net weight Tare is not 0 Indicates an ACCUMULATION transaction lb kg Current unit of measure and Pcs in PERCENTAGE and COUNTING modes HI OK LO Scale is working in COMPARISON...

Page 8: ...change a data value Selects the unit of measure opens the COMPARISON limit settings In Menu mode selects goes to next data value position Tares the scale In Menu mode opens menu option confirms a valu...

Page 9: ...em to be tared e g hand truck on the scale platform Weight of the item is displayed in the weight window 3 Press the TARE key The weight is tared 0 weight is displayed and the Net annunciator lights 4...

Page 10: ...t print out format MULTIPLE 1 Power up the scale and zero the display if necessary by pressing the ZERO ON OFF key Be sure the scale is displaying weight in your preferred unit of measure Press the UN...

Page 11: ...b 2 kg and doesn t exceed the HLD RNG oscillations range 5 div 2 5 lb 1 kg within the AVG TIM average time 5 sec the indicator calculates and displays the Hold mode weight The Hold annunciator stops f...

Page 12: ...e used to record totals of individual weighments and transactions number The DS1000 LCD is featured with two accumulation modes Manual and Automatic default The ACC mode can be disabled and the mode c...

Page 13: ...key when the COMP message pops up Press the TARE key to confirm 2 The HI annunciator lights Adjust the upper limit pressing the keys Press the TARE key 3 The LO annunciator lights Adjust the lower lim...

Page 14: ...econds Press the keys to navigate menu options 7 Select the PERCNT option and press the TARE key to confirm The annunciator flashes and the last stored item weight is recalled 8 Place the item s to be...

Page 15: ...TARE key to confirm or the ZERO ON OFF key to cancel the operation exit the menu 5 1 1 User Options USER SubMenu1 SubMenu2 Option Default Description RESET NO NO NO Returns the USER parameter to the f...

Page 16: ...GLE EH SCP SCP 12 Eh sp2 Lfuulf Lfuu Lf lf Lf uulf uu lf USB Port com2 BAUD RT 600 9600 Sets the baud rate of com2 1200 2400 4800 9600 19200 BYT FMT 8N1 8N1 Sets the bits format of com2 7O1 7E1 7O2 7E...

Page 17: ...Lf lf Lf uulf uu lf BEEP KEY YES NO YES Yes enables the beep when a key is pressed COMPAR mode NONE NONE COMPARE function modes NONE disabled L LOW IN LMT O HIGH OUT LMT HOLD Function HLD MOD NONE AU...

Page 18: ...oscillation range that can be averaged in Hold mode 0 any weight range can be averaged 1 255 only the weight which oscillates within the 1 255 div range can be averaged NLD RNG No Load Detected Range...

Page 19: ...than 7F E g If the code received is 0x31 the indicator displays 2 RD 31 232 TD RS232 send data testing the indicator displays 2 TD 55 and the serial port continuously transmits 0x55 Use the and keys...

Page 20: ...Calibration Menu The number of calibrations is shown Navigate the menu using the keys press the TARE key to confirm or the ZERO ON OFF key to cancel the operation exit the menu CAL SubMenu1 SubMenu2 O...

Page 21: ...bration Point CAL P1 and the scale capacity will be displayed 7 Use the keys to change the CAL P1 weight value 10 of the scale capacity 8 Place the certified weight over the scale platform 9 Press the...

Page 22: ...0001 50 0 5 USA 0 2 UK EU Sets the weight division 1 div 0 5 lb value of the first unit of measure The division value of the secondary unit of measure 1 div 0 2 kg is automatically determined by the i...

Page 23: ...ght type as current initial zero point when current weight is outside the IZSM range DSP OVR if the initial zero is over range display the 0 error WEIGHT current weight CAL ZRO calibration zero LAST Z...

Page 24: ...he Compar mode ACCUMU YES NO YES Yes select the Accumulation mode MANUAL add up the current value to memory when the TOTAL key is pressed AUTO add up the current value to memory when the scale is stab...

Page 25: ...Power 9V 600MA adaptor body with USA plugs included and 6xAA batteries not included Operating temperature range 32 104 F 0 40 C max relative humidity 80 at 86 F 30 C Communication interface Full dupl...

Page 26: ...ver capacity 1 at zero point 1 over capacity 2 0 RAM ok 0 ROM ok 0 gross weight 0 not in HOLD 1 RAM error 1 ROM error 1 net weight 1 in HOLD 3 0 eeprom OK 0 calibration ok 0 initial zero ok 0 battery...

Page 27: ...LF H1H2H3H4 CR ETX simulate UNIT key L CR 4c 0d LF H1H2H3H4 CR ETX simulate HOLD key X CR 58 0d power off the scale simulate OFF key others LF CR ETX EH SCP Bit Status Byte 0 0 Stable weight data 1 S...

Page 28: ...as W above others Un known commands STX Status Byte CR SCP 12 Bit Status Byte1 Status Byte2 0 0 Scale in motion 1 Under capacity 1 Stable 0 Not under capacity 1 0 Scale at zero 1 Over capacity 1 Not...

Page 29: ...LF p00xxxxxOZ CR hh ETX Scale status only if initial zero error LF hh CR ETX S CR 53 0D Read scale status LF hh CR ETX Z CR 5A 0D Simulate ZERO key no response from scale others Un known commands LF C...

Page 30: ...ll free 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 Email sales brecknellscales com http brecknellscales com Brecknell UK Foundry lane Smethwick West Midlands England B66 2LP Tel 44 0 8452 466717 F...

Page 31: ...AWT35 100107 Rev AB 31 30...

Reviews: