8
9
What to do if ...
Problem
Reason
Solution
Batteries are flat
Insert new batteries
ERR
Measurement /
handling error
Repeat measurement as
described under «Taking a
measurement». If problem
persists, contact an authorized
Service Centre.
– – –
Inflation above
320 mmHg
Repeat measurement
While measuring
(during deflation of
the cuff),
you can hear a
clicking sound with
every detected
heartbeat.
The device works
with a special
deflation procedure
(stepwise deflation).
none
Readings from
repeated
measurements
differ considerably.
Blood pressure is a
fluctuating value. For
healthy adults,
deviations of 10 to 20
mmHg are possible.
none
Cuff is not properly
fitted.
Ensure that the monitor is
fitted as follows:
– on the palm side of your
wrist
– approx. 2 cm away from the
palm line
Problem
Reason
Solution
Readings from
repeated
measurements
differ considerably.
Readings were not
consistently taken at
heart level.
For each measurement, hold
the monitor at heart level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
While measuring, relax, keep
still, do not move or speak.
Specifications
Model:
BP 1750 / BP 1700
Operation principle:
Oscillometric method
Display:
Liquid crystal display
Measuring range:
0–320 mmHg (cuff pressure)
40–199 / min (pulse rate)
Blood pressure
measurement:
20 mmHg (minimum diastolic value)
300 mmHg (maximum systolic value)
Laboratory accuracy:
± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
Clinical accuracy:
According to AAMI-SP10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Inflation:
Automatic
Deflation:
Step valve
Power supply:
2 batteries, type LR 03 (AAA)
Operating temperature:
+10 °C to +40 °C
Storage temperature:
– 20 °C to +60 °C
Humidity:
Up to 85% RH
Cuff:
Suitable for wrist circumference of 13 to 21 cm
Clinical validation:
According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial
measurements can lead to different results for devices
which are calibrated against auscultatory reference.)
If device is not used within specified temperature and humidity ranges the
technical accuracy can not be ensured.