guidance and manufacturer’s declaration – electromagnetic emissions

The IR thermometer equipment is intended for use in the electromagnetic environment specified below.   
The customer or the user of the IR thermometer should assure that it is used in such an environment.

Emissions Test

compliance

Electromagnetic environment – guidance

RF Emissions  
CISPR 11

Group 1

The ME equipment uses RF energy only for it’s internal func-
tion.  Therefore, its RF emissions are very low and are not likely 
to cause any interference in nearby electronic equipment.

RF Emissions  
CISPR 11

Class B

Complies

Harmonic emissions 
IEC 61000-3-2

Not Applicable

The ME equipment is  
solely battery powered.

Voltage fluctuations /  
flicker emissions

Not Applicable

guidance and manufacturer’s declaration – electromagnetic immunity

The IR thermometer is intended for use in the electromagnetic environment specified below.   
The customer or the user of the IR thermometer should assure that it is used in such an environment.

Immunity test

IEc60601 test 

level

compliance level

Electromagnetic environment 

– guidance

Electrostatic discharge 
(ESD)
IEC 61000-4-2

±6kV Contact
±8kwww Air

Complies

Floors should be wood, con-
crete, or ceramic tile.  If floors 
are covered with synthetic 
material, the relative humidity 
should be at least 30%

Radiated RF
IEC 61000-4-3

3 V/m 80MHz to 
2.5GHz

Complies

Field strengths outside the 
shielded location from fixed RF 
transmitters, as determined by 
an electromagnetic site survey, 
should be less than 3 V/m.
Interference may occur in the 
vicinity of equipment marked 
with the following symbol:

Separation distance calculation 
provided above.  If a known 
transmitter is present the 
specific distance can be 
calculated using the equations.

Conducted RF
IEC 61000-4-6

3Vrms 150kHz to 
80MHz

Not Applicable
(no electrical 
cabling) 

Electrical fast transient
IEC 61000-4-4

±2kV power line
±1kV I/O lines

Not Applicable

The ME equipment  
is solely battery powered. 

Surge
IEC 61000-4-5

±1kV  
differential
±2kV common

Not Applicable

Power frequency magnetic 
field
IEC 61000-4-8

3 A/m

Complies

Power frequency  
magnetic fields should be at 
levels characteristic of a  
typical location in a typical 
commercial or hospital  
environment.

Voltage dips, short          
interruptions and voltage 
variations on power  
supply input lines
IEC 61000-4-11

>95% dip 0.5    
cycle
60% dip 5 cycles
70% dip 25 
cycles
95% dip 5 sec.

Not Applicable

The ME equipment  
is solely battery  
powered. 

non-Life support Equipment separation distance calculation  

(3Vrms / 3V/m compliance)

Rated maximum output 
power of transmitter (W)

Separation distance according to frequency of transmitter (m)

150 kHz to 80 MHz 
in ISM bands

80 MHz to 
800 MHz

800 MHz to 2.5 GHz

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.17

1.17

2.33

10

3.69

3.69

7.38

100

11.67

11.67

23.33

Limited warranty

This product comes with limited warranty commencing on the date of purchase (See warranty card 

for details). Within the warranty period we will eliminate, free of charge, any defects in the appliance 

resulting from faults in materials or workmanship, by replacing the complete appliance.

This warranty is applicable only for the appliance supplied by the appointed distributor. This warranty 

does not cover: damage due to improper use, normal wear or use as well as defects that have a 

negligible effect on the value or operation of the appliance. The warranty becomes void if repairs are 

undertaken by unauthorized persons and if original Braun parts are not used.

d

 = [3.5]   

P

V

1

d

 = [3.5]   

P

E

1

d

 = [ 7

 

]   

P

E

1

Summary of Contents for IRT 6030

Page 1: ...ThermoScan Infrared ear thermometer IRT 6030 IRT 6520...

Page 2: ...rl Braun GmbH Kronberg Germany ThermoScan is a registered trademark of Kaz Europe S rl ThermoScan Kaz Europe S rl Manufactured by Kaz Europe S rl Place Chauderon 18 CH 1003 Lausanne Switzerland 2014...

Page 3: ...2 5 10 4 9 6 7 8 11 1 Warnings and Precautions Please consult your doctor if you see symptoms such as unexplained irritability vomiting diarrhea dehydration changes in appetite or activity seizure mu...

Page 4: ...ient e g liver heart lung kidney Fever may be blunted or even absent in elderly patients This thermometer contains small parts that can be swallowed or produce a choking hazard to children Always keep...

Page 5: ...position error See Errors and troubleshooting section for more information 6 7 9 6520 6030 The confirmation beep indicates that an accurate temperature measurement has been taken The result is shown o...

Page 6: ...put a new lens filter on and take a temperature measurement If the probe tip is damaged please contact the local service center Use a soft dry cloth to clean the thermometer display and exterior Do no...

Page 7: ...2 years 350 measurements Service life 5 years This thermometer is specified to operate at 1 atmospheric pressure or at altitudes with an atmospheric pressure up to 1 atmospheric pressure 700 1060hPa S...

Page 8: ...known transmitter is present the specific distance can be calculated using the equations Conducted RF IEC 61000 4 6 3Vrms 150kHz to 80MHz Not Applicable no electrical cabling Electrical fast transient...

Page 9: ...30 20 ExacTemp 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 11 10 IRT 6030 3 2 5 9 4 8 6 7 10 1 IRT 6520 3 2 5 10 4 9 6 7 8 11 1 IRT 6520 10 40 C 50 104 F 25 C 13 F 55 C 131 F 95 RH ThermoScan LF40 IRT 6520...

Page 10: ...1996 p 919 2 Guyton A C Textbook of medical physiology W B Saunders Philadelphia 1996 p 754 5 3 Netter H F Atlas of Human Anatomy Novartis Medical Education East Hanover NJ 1997 pp 63 95 2 5 AgeSmart...

Page 11: ...OS 6 7 9 6520 6030 0 3 3 36 36 Months 0 3 3 36 36 Months 0 3 3 36 36 Months 0 3 3 36 36 35 8 37 4 C 96 4 99 4 F 35 4 37 6 C 95 7 99 6 F 35 4 37 7 C 95 7 99 9 F 37 6 38 5 C 99 6 101 3 F 37 7 39 4 C 99...

Page 12: ...20 21 20 1 1 2 3 6520 15 sec 6030 15 9 Mem Mem M Mem M M2 5 Mem 5 ThermoScan C F 3 C SET F SET 1 2 ThermoScan LF40 1 5 AA LR 06 Th er m oS ca n 1 2 3 LOT LOT11614K 2014 116 3...

Page 13: ...10 40 C 50 104 F 25 55 C 13 131 F 10 95 0 1 C or F 35 C 42 C 95 F 107 6 F 0 2 C 0 4 F 0 3 C 0 5 F 0 14 C 0 26 F 2 350 5 1 1 700 1060 hPa BF 10 C 40 C EN 12470 5 2003 EN 60601 1 2006 EN ISO 14971 2012...

Page 14: ...1000 4 6 3Vrms 150kHz 80MHz IEC 61000 4 4 2kV 1kV I O IEC 61000 4 5 1kV 2kV IEC 61000 4 8 3 A m IEC 61000 4 11 95 0 5 60 5 70 25 95 5 3Vrms 3V m W m 150 kHz 80 MHz ISM 80 MHz 800 MHz 800 MHz 2 5 GHz 0...

Page 15: ...26...

Reviews: