background image

 

 

 

Brainlab AG 

Olof-Palme-Str. 9 

81829 Munich 

Germany

 

Customer Support: 

[email protected] 

North/South America: +1 800 597 5911 

Europe & rest of the world: +49 89 991568 1044

 

Page 1/10 

a

 

BL-IL-60961-45ML Rev. 2.0 

*60961-45ML* 

Date of issue: 2021-08-18 

 

STI Robotics 

49702

 

 

Safety Technical Inspection 

ENGLISH 

 

1.

 

General 

The device shall be maintained and serviced on a regular basis, 
at least once a year, based on the requirements of IEC62353 - 

Medical electrical equipment - Recurrent test and test after 

repair of medical electrical equipment. 

For details about performance of the inspection as well as the 

assessment of the results of the tests strictly follow the 
requirements of IEC62353.  

The following sections refer to referenced chapters in the 
above mentioned standard and serve to support the inspection 
accordingly. 

Use only certified testing equipment intended and designed to 
perform tests according IEC62353. 

2.

 

Visual inspection (5.2 – IEC62353) 

-

 

The device fuse inserted at the rear end of 

Robotics

 

complies with the data given in the 

Hardware User Guide, 

Patient Support System

 (check imprinting on the fuse). 

-

 

All Warning Labels as defined in the 

Hardware User 

Guide, Patient Support System

 are complete, legible and 

show no damage. 

-

 

The 

Robotics

 housing, including the bellows is not 

damaged or loose. All parts are securely fastened. 

-

 

The covers and mechanical parts of the ON/OFF-

switches are not damaged, loose or otherwise impaired 
by wear. 

-

 

The covers and mechanical parts of the Lock-All push 

buttons are not damaged, loose or otherwise impaired 
by wear. 

-

 

The cover of the Battery Charger does not show any 
cracks or other damage. 

-

 

No twists or cracks in cable jacket and no damage to  
connectors 

Open the left cover shell of 

Robotics

. Inspect the 

functional earth conductor (FE) connecting the upper 
and lower metal housing plate and verify the 

connection is not impaired by wear or fatigue (see 
figures 1/2/3 for details): 

-

 

There are no obvious cracks in the isolation or sharp 
bending of the conductor 

-

 

The mechanical connection of the conductor to upper 

AND

 lower housing plate is not loose 

3.

 

Measuring of protective earth (PE) resistance (5.3.2 – 
IEC62353) 

Disconnect the Battery Charger and switch 

Robotics

 ‘OFF’. 

Verify that the resistance of the protective earth connection 
between upper and lower metal housing plate of 

Robotics

 

complies with the requirements of IEC62353 (see figure 4 for 
details). 

-

 

The measured internal PE-resistance shall not exceed 
0.2 

4.

 

Leakage currents (5.3.3 – IEC62353) 

Perform an Equipment Leakage Current test with 

Robotics

 

and 

verify the Equipment Leakage Current from accessible 

conductive parts complies with the requirements of IEC62353 
(see figure 5 for details). 

-

 

The Equipment Leakage Current shall not exceed the 
limits as specified in IEC62353 

-

 

Allowed limits differ depending on the type of 

measurement setup. For details please refer to 
IEC62353 

5.

 

Measurement of insulation resistance (5.3.4 – IEC62353) 

The insulation resistance shall only be measured at the Battery 

Charger. Therefore disconnect the Battery Charger from 

Robotics

Verify the insulation resistance between the primary and 
secondary circuits of the Battery Charger in accordance with 
the requirements of IEC62353 (see figure 6 for details). 

-

 

The insulation resistance shall meet as a minimum the 
limits as specified in IEC62353 

6.

 

Functional test (5.4 – IEC62353) 

The safety related functions of the equipment shall be tested 
as follows: 

-

 

Test the Service Lock mechanism: 

Both arresting bolts can be operated easily to both the 

‘Locked’ and the ‘Unlocked’ position. 
The indicator lights in the ‘Lock All’ push buttons are lit 

on both sides of Robotics if an arresting bolt is not in 
its normal ‘Unlocked’ position. 

-

 

Test the Service Brake Release: 
Follow the instructions in the 

Hardware User Guide, 

Patient Support System 

to perform a Service Brake 

Release. 

The function works as described. After the test 

Robotics indicates the occurrence of a Service Brake 
Release and a ‘Homing’ procedure is required. 

-

 

Inspect and test the correct function of the ON/OFF-

switches: 

The device can be turned ‘ON’ and ‘OFF’ by operating 
the switches on either side. Both switches are lit to 
indicate the ‘ON’-state of 

Robotics

 

-

 

Inspect and test the correct function of the Lock-All 
push buttons: 

The device can be ‘Locked’ and ‘Unlocked’ by 

operating the push buttons on either side. Both buttons 
are lit to indicate the ‘Locked’-state of 

Robotics

 

-

 

Test the Emergency OFF (EMO): 
 

 

-

 

Danger of damage to the Linac system if the 
Emergency Off switch is activated while the Linac is 
in operation! 
Contact local service specialist to attend and 
supervise the testing!

 

 
Perform a

 Robotics

 maneuver. During movement press 

the Emergency OFF push button of the Linac:

 Robotics

 

stops immediately. 

7.

 

Reporting of results (6.1 – IEC62353) 

All tests performed as stated above shall be documented 
comprehensively. 

For further information please see the 

Hardware User Guide, 

Patient Support System

 or consult Brainlab customer support.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

figure 1: 
Opening left cover shell

 

 

figure 2: 
No obvious cracks in the isolation or sharp bending of the 
conductor

 

  

figure 3: 
The mechanical connection of the conductor to upper and 
lower housing plate is not loose

 

 

 

 

Summary of Contents for STI Robotics 49702

Page 1: ...53 Perform an Equipment Leakage Current test with Robotics and verify the Equipment Leakage Current from accessible conductive parts complies with the requirements of IEC62353 see figure 5 for details...

Page 2: ...I Robotics 49702 Safety Technical Inspection figure 4 Reference points for PE resistance test figure 5 Reference points for leakage current test figure 6 Reference points for insulation resistance tes...

Page 3: ...ene interne PE Widerstand darf nicht ber 0 2 liegen 4 Ableitstrom 5 3 3 IEC62353 F hren Sie einen Ableitstromtest mit Robotics durch und stellen Sie sicher dass der Ger teableitstrom von zug nglichen...

Page 4: ...STI Robotics 49702 Sicherheitstechnische Pr fung Abbildung 4 Referenzpunkte f r PE Widerstandstest Abbildung 5 Referenzpunkte f r Ableitstromtest Abbildung 6 Referenzpunkte f r Isolationswiderstandst...

Page 5: ...3 IEC62353 Eseguire un test della corrente di dispersione dell apparecchiatura Robotics e verificare che la corrente di dispersione sia conforme ai requisiti previsti dalla norma IEC62353 per ulterior...

Page 6: ...rimento per il test di resistenza di messa a terra di protezione Figura 5 Punti di riferimento per il test della corrente di dispersione Figura 6 Punti di riferimento per il test di resistenza dell is...

Page 7: ...a con Robotics y compruebe que la corriente de fuga procedente de los componentes conductores accesibles cumple los requisitos de la norma IEC62353 ver la figura 5 La corriente de fuga del equipo no d...

Page 8: ...dad Figura 4 Puntos de referencia para la resistencia de la puesta a tierra Figura 5 Puntos de referencia para la prueba de corriente de fuga Figura 6 Puntos de referencia para la resistencia de aisla...

Page 9: ...ue le courant de fuite de l quipement par les pi ces conductrices accessibles r pond aux exigences de la norme CEI 62353 voir la figure 5 pour plus de d tails Le courant de fuite de l quipement ne doi...

Page 10: ...t figure 4 points de r f rence pour le test de r sistance du conducteur la terre figure 5 points de r f rence pour le test du courant de fuite figure 6 points de r f rence pour le test de r sistance d...

Reviews: