![Bowket HL-A04 Operation Manual Download Page 12](http://html.mh-extra.com/html/bowket/hl-a04/hl-a04_operation-manual_2798033012.webp)
Shenzhen Bowei tech Co.ltd.
ADD:RM902;ShangMei Times;NO.57;LongGuan RD;Long Hua Dist Shenzhen, China;518000
Tel:86-755-29994192 13148703639
Email:[email protected] http://www.bowei-ch.com/
Water flow status:
The icon
represents the normal recycling water in the system while
represents the abnormal recycling water in the system. When the water flow is
blocked or there is no water in the system, the icon will alternately display and
give off the alarm sound at the same time. At this time, all the keys on the
screen could not be used.
Bareheaded status:
The icon
represents that the bare head has been lit; the icon
represents that the bare head has not been lit. After the bare head is lit, it could
enter into the charged status. If it does not light the bare head for five seconds,
then return to the ready mode.
Charged status:
The icon
represents successful charge while the icon
represents
unsuccessful charge.Only after the successful charge could it enter into the
treatment status for treatment. If the charge is not successful for five seconds,
then return to the ready mode.
OPERTATION STEP
1) Fill water to machine, until water is out of overflow, it means the water is full
of water tank.
2) Choose treatment head: 1064nm, 532nm.Then fix it in the laser gun.
3) Connect the power line to machine, turn on the urgency stop button and key
switch, the screen will be light, you will hear sound from the machine, it is