Safety
Basic Construction:
In accordance with EN 60601-1
Mode of Operation:
Intermittent operation
Protection Class:
CLASS I EQUIPMENT
Output Type:
TYPE BF
Dimensions and Weight
Length x Width x Height = 9" x 4.5" x 6.3"
Weight: 5 lbs.
WARRANTY AND REPAIR
The Aaron High Frequency Desiccator is covered under warranty for a period of two years. The handpiece is covered
under warranty for a period of one year or 25 steam autoclave cycles, whichever comes first. The warranty becomes
null and void if damage occurs from incorrect handling or misuse of the product.
For warranty and repair work, please contact Aaron and obtain a Return Goods Authorization number (RGA). Place
the number so that it can be seen on the exterior of the package and ship directly to Aaron. A return without an
RGA may not be accepted.
TROUBLESHOOTING
The Aaron High Frequcncy Desiccator has been designed and manufactured with the utmost safety in mind. The
unit is equipped to automatically detect a malfunction. The following table list error codes, their meaning and
recommended actions to be taken to resolve the error.
Error Code
Description of Error
Recommended Action
E1
Activation upon power up
• Check handpiece for activation.
• Check footswitch for activation; once the activation is
halted the unit will resolve the error. If the error
persists the handpiece could be malfunctioning and
may need to be replaced.
E2
DC Supply Over Voltage
• Switch unit off and on again.
Detection
• Make sure unit is connected to correct power source
for the unit.
E3
Pulse Width
• Switch unit off and on again.
E4
Delta Error
• Switch unit off and on again.
E5
Multiple Errors
• Switch unit off and on again.
E6
Temperature Error
• Switch unit off. Allow unit to cool. Switch unit on.
If problems persist, the unit should be taken out of service and the manufacturer should be notified. For technical
support or return authorization phone +1-800-537-2790.
TRANSPORT AND STORAGE
Ambient temperature range: -40˚C to 70˚C
Relative humidity: 10% to 100%, non-condensing
Atmospheric pressure: 500hPa to 1060hPa.
8
Bovie Medical
Summary of Contents for AARON 800-EU
Page 1: ...1 User s Guide Aaron 800 EU 900 USER S GUIDE AARON 800 EU AARON 900...
Page 11: ...GUIDE DE L UTILISATEUR 11 AARON 800 EU AARON 900 User s Guide Aaron 800 EU 900...
Page 21: ...21 BEDIENUNGSANLEITUNG AARON 800 EU AARON 900 User s Guide Aaron 800 EU 900...
Page 31: ...31 User s Guide Aaron 800 EU 900 GUIDA PER L UTENTE AARON 800 EU AARON 900...
Page 41: ...41 User s Guide Aaron 800 EU 900 GU A DEL USUARIO AARON 800 EU AARON 900...
Page 51: ...51 User s Guide Aaron 800 EU 900 GUIA DO UTILIZADOR AARON 800 EU AARON 900...