7.5.2 Trouble shooting handheld
Display
Description
Solution
Regeneration is ongoing.
Wait for the purging to be completed. Do not remove the
mouthpiece.
Battery is almost empty.
Charge the handheld on the basestation cradle, see
“3.1 Installing the handheld”.
Disposable mouthpiece is not valid.
Exchange the mouthpiece and take a new one from the package,
see “4.1 Measurement using the basestation”.
Measurement procedure is incorrect,
breathing was too strong, too weak or
time-limit after regeneration was
exceeded.
Repeat the measurement, see “4.2 Measurement stand-alone
with the handheld”.
Permissible number of measurement
trials reached soon, measurements still
possible.
See display for number of remaining measurement trials. Order
new
Vivatmo pro
handheld, see “7.2 Status information hand-
held”.
Number of permissible measurement
trials reached. Device does not perform
any measurements after maximum
number of permissible tests is reached.
Connect a new
Vivatmo pro
handheld and dispose of the old
one, see chapter “8.1 Disposing the
Vivatmo pro
device and the
battery”.