6. In accordance to IT definitions in your organization, define the location for the
inbox for incoming measurement orders.
7. Repeat the procedure for the location of the outbox.
8. Confirm with
DONE
.
When credentials are required, follow steps 9 to 13.
Username
Enter Username …
Order/Result Interface
Z
X
C
V
B
N
M
,
.
. ? 1 2 3
D o n e
A
S
D
F
G
H
J
K
L
W
E
R
T
Y
U
I
O
P
Q
9. Enter the user name required to access the data from the EPR system.
10. Confirm with
DONE
to move to the next screen.
11. Enter the password required to access the data from the EPR system.
12. Confirm with
DONE
.
Order/Result Interface
GDT
Protocol
smb://example/vivatmo/inbox
Inbox Location
smb://example/vivatmo/outbox
Outbox Location
active
Order/Result Interface
Praxis-admin-1
Username
••••••••••
Password
Positive
Communication Test
S A V E
The settings for the
ORDER/RESULT INTERFACE
are displayed including the
result of a Communication Test.
13. Check the data and confirm with
SAVE
when the data are correct.
On your home screen, you see the order button, see “2.1 Screen elements”.
If you delete the inbox and the outbox, the order button disappears from the
home screen.