20
en | Installation
Carephone 62
F.01U.141.162 | V6 | 2015.04
User Manual
TeleAlarm
The GSM Module will initialize. An LED indicator is visible.
See
Section 7.6.5 GSM Module LED indicator, page 47
.
Close the cover & insert the power plug in the Carephone.
Step 4: test the GSM Module
1.
Test the signal strength: see
Section 7.6.3 Test the GSM
Module signal strength, page 46
.
2.
Perform a test call:
see Section 6.3 Test mode, page 38
.
5.3.5
Connection configuration
To check whether an emergency call can be sent:
1.
Make a phone call as you normally would.
2.
When the phone call is in progress, press the red button.
If the setup is correct, the phone call will be interrupted.
5.4
Deinstallation
To switch it off, remove the power plug from the Carephone.
LED
NOTICE!
In the event of an emergency, the unit must be able to send out
an emergency call even if a phone call is in progress at the time.
CAUTION!
With a private branch exchange (PABX) there is no guarantee
that a phone call will be interrupted by the emergency call.
NOTICE!
If you remove the power plug from your power outlet, the unit
announces "
power failure"
and switches to battery operation.