19
or over 10% of full capacity.
( In case of less 1/5,000 resolution ,the value of standard weight
must be over 10% of full capacity at least.)
( In case of over 1/5,000 resolution ,the value of standard weight
must be over 20% of full capacity at least.)
(Notice) If span capacity is set less 10% , indicator will display
error message.
Key
Display
Description
“ENTER” key :
Store and move into
next menu
Step 3
100 kg
10000kg
Setting Weight
☞
REF
1. The weight shall be within the range of 10%~100% of maximum weight.
☞
REF
2. The setting weight must be over the range of 10% of maximum weight.
☞
REF
3. The setting weight over the maximum capacity.
GhYd(
-
Function :
NYfc7U`]VfUh]cb
A step to check the zero conditions of Indicator.
After appearing "UnLoAd", please push ENTER key.
Please do it as the zero adjustment instruction.
Key
Display
Description
“ENTER” key :
Zero calibration
and move into
next menu
Step 4
Unload the tray and press “ENTER”
key Under zero calibration
Zero calibration is completed.
Summary of Contents for BS-8300
Page 1: ......
Page 8: ...7 8 A9BG CBG ...