▣ What is Simulation Calibration?
Calibration is performed for matching the BS-8300 to a strain-gage
sensor. The following two type of calibration are available for the BS-
8300.
%"G]ai`Uh]cb58>IGHA9BH
▣ Equivalent input Calibration.
Calibration is performed without an actual load by entering the rated
output value (mV/V) and the capacity (to be indicated) of the strain-
gage sensor by the keys. Calibration is easily performed when no actual
load is available.
For example, the gain is automatically determined by entering.
2.001mV/V (rated output)-100.0kg (capacity)
as indicated for a load.
%!% <ckhcUWWYggh\YGD5B58>IGHA9BH
Turn on the power while
pushing Key then, the display will
be "Sim. CAL."
Also, pushing ENTER KEY on the below right. it will be displayed with "d xx"
("xx" means 1, 2, 5, 10, 20, 50, 100)
예) POWER OFF CONDITIONS
1. While pushing
e
Key ---------- Display is "Sim. CAL."
2. Pushing ENTER key ---------- Display is "d 2"
☞ "Sim. CAL." means SETUP & CALIBRATION mode
%!& 5jU]`UV`Y?Ymg
S&C MODE have 6way to adjust span. each step will be advanced with
ENTER key. Also, ESC key was used to return the prior conditions.
※ F.F : ENTER key
※ Review : ESC key