63
15. Repair
15-1 Error message
In case of error display, please take measure accordingly with the countermeasures.
Error display
Cause
Countermeasure
Key lock function is operated.
Conduct key lock cancellation with the use of
14-1 KEY Lock cancellation method.
1. Dead weight value is small.
2. Voltage of span calibration
conducted point is either same or
lower than zero point.
3. There is no change in load cell
output or change upon small dead
weight value.
This error is occurred under calibration status.
Calibration shall be conducted using appropriate
dead weight.
Check the disconnection of load cell since (-)
output of load cell is conducted.
Load cell output value is high.
This error is occurred when checking load cell
output voltage of check mode.
Abnormality of load cell shall be checked since
load cell output is 35mV or higher.
It shall be less than 15mV or lower in general.
Load cell output value is low.
This error is occurred when checking load cell
output voltage of check mode.
Abnormality of load cell shall be checked since
load cell output is -35mV or lower.
It exceeded max range of display value.
It exceeded max range of display that is either -
199999 or 999999.
Conduct calibration again after checking load cell
output voltage and disconnection.
15-2 Load cell inspection
When the load cell in use is instable, following items shall be inspected.
1.
Check whether or not load cell terminal is properly connected to the meter.
2.
Check whether or not terminal connection within Summing Box and Junction Box is properly
conducted.
3.
Check whether or not there is mechanical interference.
4.
When the zero point value of load cell, namely output value under no load status of load cell,
strays off from self-specification range, it may be caused by mechanical transformation by
overloading and impact, interference of meter container and structure, damage in inner circuit
of load cell, etc.
5.
Check whether or not measured resistance between terminals of load cell conforms to the
specification (refer to catalogue and report).
6.
Measure the insulation resistance between each lead wire (wire in red, white, green, and blue)
and earth wire with 50V D.C insulation tester and check whether or not it is higher than 100 ㏁.