6-5-3 Zero Tracking
Automatic update of zero point shall be conducted by detecting the movement of zero point with the use
of zero tracking function (setting shall be conducted at function).
Zero tracking shall only be conducted when width and time function of zero tracking are already set.
Zero tracking is not conducted when zero point is within zero calibration range.
Zero Calibration
Width of Zero Tracking
nonE None
Width of zero tracking shall be set.
0.000 Setting Range
0 ~ 999
Time of Zero Tracking
00 Setting Range
0 ~ 99
Zero tracking operation time shall be set.
00: Operation within 0.,1 sec.
01: 0.1 sec
~
99: 9.9 sec
Warning: When ‘None’ is selected at zero
tracking width setting, zero tracking time
setting mode shall not be displayed.
Display 0
Tracking width
Start
Tracking time
Zero
calibration
cannot
be
conducted
under
Key
Lock
status
thus
it
shall
be
executed
after
cancelling
Key
Lock
status.
Zero
tracking
is
not
operated
when
offset
value
is
set
at
function.
When
there
is
a
dramatic
change
in
load
by
vibration
even
within
the
range
of
zero
tracking
width,
zero
tracking
setting
times
is
doubled
thus
there
may
be
a
delay
in
conduct
of
automatic
zero
calibration.