13-1-5
BCD output check
It is a mode to check BCD output.
BCD of associated digit value is output accordingly with the input value of meter display unit.
For instance,
100
digit is output with the input of 000001 and 000002. 111111, 222222, 234985……
It is a mode to check BCD
output. Device to use shall be
checked through connection.
Output of changed value is
conducted from BCD output
terminal when there is a changed
in value with the use of No.2 and
No.3 key.
It moves onto next stage
when pushing No.4 key and
push No.1 key for
cancellation.
Check
mode
is
displayed
only
when
BCD
is
selected
from
13-1-3 Optional output (DAC, BCD)
selection mode
.
For
range
of
setting
value
change,
setting
is
available
within
‐
199999
~
999999.
Decimal
point
and
(
‐
)
output
are
also
available.
(initial
value
is
set
as
000.000)
13-1-6 Initialization Mode
Various settings of data are initialized.
All settings including calibration, function setting, optional output setting, and others are initialized.
Take caution since calibration setting is also initialized.
In case of conducting actual load calibration, dead weight may be necessary upon recalibration.
Select whether or not to conduct
initialization with the use of No.2
and No.3 key. Select Yes for
initialization and No for no
initialization.
Initialization is conducted when
pushing No.4 key under Yes
status.
No initialization is conducted
when Pushing No.1 key.
Check mode has been
completed and it is turned
into measurement mode.
Entry
to
check
mode
is
available
even
under
key
lock
status.
Actual
load
calibration
is
initialized
and
it
returns
to
factory
forwarding
status.