10 Technical Data
bon SL-85
10 Technical Data
Slit lamp:
Classification (EN 60601-1)
II, Type B
Slit lamp measurements (H/W):
500x380 mm
Weight: 9.5
kg
(excl.
mains
adaptor)
Slit projector halogen lamp:
6V/20W PG 22 (Standard)
Focussing
light
lamp:
12V/60mA
Slit length (continuous):
1.8 -12 mm
Slit
width
(continuous): 0-14
mm
Smallest
point: 0.3
mm
Maximum
slit
length:
14
mm
Measurements of the light field:
(0.3/5.5/9/14) mm
Filter:
blue,
green,
yellow,
grey
Rotation
of
the
slit:
+/-
90°
Movement of the slit lamp to/from the microscope: +/- 90°
Magnification:
6x/10x/16x/25x/40x
Ametropia
equalisation:
+/-
5
dpt
Base (x,y,z):
Sideways adjustment (left/right):
108 mm (x)
Lengthways adjustment (forwards/backwards):
113 mm (y)
Height
adjustment:
approx.
40
mm
(z)
Mains adaptor:
Input supply voltage:
100-120V AC / 230-240V AC
Output voltage:
6V AC (slit lamp), 12V AC (focussing light)
Power frequency:
50/60 Hz
Connection power:
40VA
Fuses:
2 x 0.4A T (100-120V) / 2 x 0.2A T (230-240V)
Safety class:
II
Safety type:
IP 21
Device type:
B
Label Symbols
Transport requirements
Protective earth
Temperature: -5 °C to +45 °C (+23 °F to +113 °F)
Safety class II (isolated)
Air pressure: 650 hPa to 1100 hPa
Fuse
Relative humidity: 25% to 80%
Read instruction manual
Application part type B
Maximum conditions – no longer than 60 days in a row
GA bon SL-85 Rev 1.0 E 211005.doc
18/20