7  Operation of the apparatus 

 

 

 

bon SL-85 

7  Operation of the apparatus 

7.1  Focussing the microscope 

Remove the cap and place the test rod in the holder  
as shown. Turn the slit lamp on. 
Focus the eyepiece so that the grainy surface  
of the test rod can be seen in focus at  
medium magnification. 
 
 
 
 
 
 
 

                                                                                                      Diagram 7-1:Slit projector 

 

7.2 Examination 

Š

  Allow the patient to rest his/her chin comfortably on the chin rest and ensure that 

his/her forehead is on the forehead rest.  

Š

  Use the height adjuster to bring the eyes of the patient up to the level of the marking 

on the head rest (line of vision height). 

Š

  Turn the slit lamp on. 

Š

  Look through the microscope and set the desired magnification. 

Š

  Centre and focus the eye to be examined with the control lever on the base. 

Š

  In order to reduce patients‘ discomfort, the brightness can be adjusted using the 

mains adaptor. 

 
When directly lighting the eye, the slit of the slit projectors is to be projected in line with the 
viewing level of the microscope. The height and width of the slit can be adjusted on the slit 
projector (see diag. 4.3, page 9). With the slit made smaller, the ray of light from the slit lights 
a small section of the eye, placing it into high contrast. When widened, the section of the eye 
that can be examined is larger, but the contrast is not so high. 
 
For examination in a red-free image, a green filter is available. Turn the filter dial to the 
position with the green markings. The blue filter aids observation of intraocular pressure 
measurements, e.g. with an Applanation tonometer. The grey filter can be used to protect the 
patient’s eye from heat. The yellow filter on the microscope (see diag. 4.2, page 8) can be 
used in conjunction with the blue filter for fluorescence observation, e.g. to view the position 
of contact lenses.       

GA bon SL-85 Rev 1.0 E 211005.doc

 

 

13/20

 

 

 

Summary of Contents for SL-85

Page 1: ...Internet www bon de Sparkasse zu L beck Sort code 230 501 01 Acc No 1 014 885 Swift BIC HSHN DE H1 SPL IBAN DE 2305 0101 0001 0148 85 Commerzbank L beck Sort code 230 400 22 Acc No 0 107 755 Postbank...

Page 2: ...tor 9 5 Setting up assembly and repair 10 6 Powering up 12 7 Operation of the apparatus 13 7 1 Focussing the microscope 13 7 2 Examination 13 8 Maintenance and care 14 8 1 Care 14 8 2 Maintenance 14 8...

Page 3: ...ctions Caution Please observe safety instructions with this symbol to prevent personal danger or damage to property Important Indicates particularly important information to maintain the function of t...

Page 4: ...Liability The slit lamp is manufactured according to the current technical status and the recognised safety regulations and is tested in accordance with strict quality criteria bon Optic only accepts...

Page 5: ...0W 1 x non slip surface 1 x mouth cover with fixing screw 1 x test rod 2 x gearwheel plates 2 x running wheel covers 1 x set of screws 1 x power socket with front cover Components needed for operation...

Page 6: ...atures of between 10 C and 40 C The ophthalmometer must not be used in areas where there is a danger of explosion Avoid dropping or splashing water on the device The slit lamp must only be used by aut...

Page 7: ...7 Moveable arm 6 Lamp housing 5 Slit projector 4 Mouth cover 8 Joystick with height adjuster 10 Axis 9 Base Diagram 4 1 Slit lamp bon SL 85 The slit lamp sits on a base 9 with orthogonal movement The...

Page 8: ...85 The eyepiece 4 with standard 12 5x magnification enables an ametropia equalisation of 5 dpt By extending or pushing in the eyepiece lever 5 the distance to the pupil can be adjusted Magnification a...

Page 9: ...for slit projector 5 Filter adjuster Diagram 4 3 Slit lamp The slit projector offers a wide spectrum of different types of lighting The slit width 3 can be continuously adjusted from 0 to 14mm The sl...

Page 10: ...lace the base on the table in such a way so that the joystick is situated over the non stick surface and the gearwheels sit on the gearwheel plate e Place one protective cover to the side of the assem...

Page 11: ...5 Setting up assembly and repair bon SL 85 Diagram 5 2 Assembly sketch SL 85 GA bon SL 85 Rev 1 0 E 211005 doc 11 20...

Page 12: ...the slit lamp into connection 3 2 Plug the focussing light into connection 2 optional 3 Earth the slit limp to connection 4 4 Plug the mains adaptor 1 into the mains socket using the power cable 5 Pr...

Page 13: ...can be adjusted using the mains adaptor When directly lighting the eye the slit of the slit projectors is to be projected in line with the viewing level of the microscope The height and width of the s...

Page 14: ...use disinfection alcohol 8 2 Maintenance When used properly the slit lamp should not require repeated maintenance For repairs or technical problems please contact the bon Optic customer services depar...

Page 15: ...the lamp to cool for several minutes before changing it 3 Open the cap 2 4 Move the clamp to the side and remove the lamp 1 in its holder 5 Place the new lamp in the holder Do not touch the glass body...

Page 16: ...ect from the mains before changing the fuse 2 Open the required fuse element 3 Remove the old fuse from the holder and replace it with a new fuse of the same type 4 Place the fuse holder back in the f...

Page 17: ...used properly and in accordance with the conditions of use Repairs have not been carried out by anyone other than the bon Optic customer service team or persons authorised by bon Optic Guarantee serv...

Page 18: ...x Ametropia equalisation 5 dpt Base x y z Sideways adjustment left right 108 mm x Lengthways adjustment forwards backwards 113 mm y Height adjustment approx 40 mm z Mains adaptor Input supply voltage...

Page 19: ...15 Mains adaptor Transformer 16 Holder for test rod 17 Lamp not visible 18 Running wheel cover 19 Teflon non slip surface 20 Screw for slit projector 21 Scale for slit projector 22 Screw for moveable...

Page 20: ...tlinie 93 42 EWG Anhang IX Regel 1 1 MDD 93 42 EEC appendix IX rule 1 Wir erkl ren hiermit die bereinstimmung des vorgenannten Produkts mit der EU Richtlinie 93 42 EWG ber Medizinprodukte We declare t...

Reviews: