Maintenance
Maintenance UK
It is a legal requirement, under regulation 9 of the COSHH
regulations that all local exhaust ventilation systems are
thoroughly examined and tested at least once every 14
months (typically carried out annually). The approved code
of practice recommends that a visual check should be
carried out at least once a week.
COSHH requires the annual inspection and testing to be
carried out by a competent person and specifies that
documentation results are recorded in a log.
Contact the seller for more information about inspection
and certification.
Maintenance General
User maintenance is limited to cleaning the unit and filter
replacement, only the manufacturers trained maintenance
technicians are authorised to carry out component testing
and replacement. Unauthorised work or the use of
unauthorised replacement filters may result in a potentially
dangerous situation and/or damage to the extractor unit
and will invalidate the manufacturer’s warranty.
Cleaning the unit
The powder coat finished units can be cleaned with a damp
cloth and non-aggressive detergent, do not use an abrasive
cleaning product as this will damage the finish.
The cooling inlets and outlets should be cleaned once a year
to prevent build-up of dust and overheating of the unit.
Replacing the Filters
The filters will need changing when the Green LED on the
front of the unit starts to flash Red.
At this point the Pre filter bag should be changed, if the
Green LED has not returned then the HEPA filter should also
be changed.
A log of the changes should be maintained by the user.
The filters require attention when the LED flashes RED or
when the extractor no longer removes fume efficiently.
It is recommended that a spare set of filters are kept on site
to avoid prolonged unit unavailability. Part numbers for
replacement filters can be found on the filters fitted in your
system. Alternatively, refer to the spare parts table in this
manual.
To prevent overheating, units should not be run with a
blocked filter condition, or with dust obstruction of Inlets /
Outlets.
Pre Filter Replacement
To replace the Pre filter follow the procedure detailed
below.
Always use Goggles Gloves and Masks when changing the
filters.
1.
Isolate the electrical supply to the extractor
2.
Undo the 2 latches on the front of the extractor to
open the front door.
3.
Pull the filter tray out of the unit and remove the
bag filter.
4.
Once removed it is recommended that the filters
are bagged and sealed
5.
Slide the new filter into place on the tray and install
the tray
6.
Close the door and secure the 2 latches
7.
Reconnect the mains supply to the unit.
01
6