4

Installation Instructions

Your balance is designed to provide reliable weighing results
under normal ambient conditions.  When choosing a location
to set up your balance, observe the following so that you will
be able to work with added speed and accuracy:

– Set up the balance on a stable, even surface
– Avoid placing the balance in close proximity to a heater or

otherwise exposing the balance to heat or direct sunlight

– Protect the balance from drafts that come from open 

windows or doors

– Avoid exposing the balance to extreme vibrations during

weighing

– Protect the balance from aggressive chemical vapors
– Do not expose the balance to extreme moisture over long

periods

Conditioning the Balance: 
Moisture in the air can condense on the surfaces of a cold
balance whenever it is brought into a substantially warmer
place.  If you transfer the balance to a warmer area, make
sure to condition it for about 2 hours at room temperature,
leaving it unplugged from AC power.

Setting up the Balance

Balances with an Analytical Draft
Shield Chamber:

§

Place the components listed
below inside the chamber in
the order given:

– Shield plate
– Shield ring
– Pan support
– Weighing pan

Balances with a Glass 
Draft Shield:

§

Place the components listed
below on the balance in the
order given:

– Shield ring – place it on the

balance so that the edge for
fitting the glass draft shield
faces upwards and turn it until
it is firmly in place.

– Pan support
– Weighing pan
– Glass draft shield
– Draft shield cover

Balances with a Round Weighing
Pan Ø 115 mm

§

Place the components 
listed below on the balance 
in the order given:

– Pan support
– Weighing pan

Balances with a Round Weighing
Pan Ø 150 mm

§

Place the weighing pan 
on the balance

Connecting the Balance to AC Power/Safety Precautions

Use only original AC adapters:

– for Europe: 6971948
– for the US:

6971947

– for the RSA: 6971949
– for Australia: 6971950

§

Insert the right-angle plug into
the jack

§

The AC adapter rated to 
Class 2 can be plugged into
any wall outlet without 
requiring any additional safety
precautions

The ground is connected to the balance housing, which can
be additionally grounded for operation.

Using a (Rechargeable) Battery

(not for models BBL31, BBL32, BBL42 or BBL52)

$

Battery is not included with
the equipment supplied

!

Use only a commercially 
available (rechargeable) 
9-volt battery

!

When using a rechargeable
battery, always use an 
external recharging device 
to recharge the battery

§

Lay the balance on its side

§

Open the battery compart-
ment:  Lift the compartment
cover

§

Insert the 9-volt 
(rechargeable) battery in 
the compartment

$

Make sure to connect 
the positive and negative
poles correctly

!

All used batteries are 
classified as waste that
requires special handling 
(not “household” waste). 
Dispose of rechargeable 
batteries in accordance with
the applicable special waste
disposal regulations.

§

Close the battery compart-
ment:  Press down on the
cover until it clicks into place

Summary of Contents for BBL series

Page 1: ...BOECO 98648 012 04 BBL Models Electronic Precision Balances and Gold Scales Operating Instructions Modelos BBL Balanzas electr nicas de precisi n para metales preciosos Manual de instrucciones...

Page 2: ...2 English page 3 Espa ol p gina 13...

Page 3: ...ee operation of your balance Make sure that the voltage rating printed on the AC adapter is identical to your local line voltage Use only a commercially available 9 volt battery or rechargeable batter...

Page 4: ...aft shield faces upwards and turn it until it is firmly in place Pan support Weighing pan Glass draft shield Draft shield cover Balances with a Round Weighing Pan 115 mm Place the components listed be...

Page 5: ...isplay 1 Turn on the balance ON OFF Self test is performed 2 Place container on balance 52 0 g here 52 g 3 Tare the balance t 0 0 g 4 Place sample in container 150 2 g on balance here 150 2 g Calibrat...

Page 6: ...2 1 4 Reference sample quantity 20 pcs 3 3 3 Step Key or instruction Display 1 Turn on the balance ON OFF 2 Tare the balance t 0 0 g 3 Display the reference FUNCTION ref 20 briefly sample quantity 2 s...

Page 7: ...ng operations or subweighs Available Features The measured result displayed is the arithmetic mean shown in the selected weight unit a triangle under the plus or minus sign indicates that this is a ca...

Page 8: ...two weight units Configure the Toggle weight units application in the operating menu See Configuration menu code 2 1 2 Toggle weight units Menu code Unit Conversion factor Abbr on printout 1 7 2 o 3 1...

Page 9: ...ctory setting User setting Menu 1 Balance 1 1 Adapt filter 1 1 1 Minimum vibration functions 1 1 2 o Normal vibration 1 1 3 Strong vibration 1 1 4 Extreme vibration 1 3 Stability range 1 3 1 1 4 digit...

Page 10: ...ro is displayed not met e g balance not zeroed Press t to tare the balance balance is loaded Unload the balance e 09 When gross zero no tare Tare the balance e 10 The t key is blocked Press CAL CF 2 s...

Page 11: ...urn off the power and disconnect the equipment from AC power immediately Lock the equipment in a secure place to ensure that it cannot be used for the time being Safe operation of the balance with the...

Page 12: ...1 0 1 0 1 0 1 1 Linearity g 0 2 0 2 0 2 0 2 2 Sensitivity drift within 10 30 C 50 86 F K 0 5 10 5 1 10 5 2 10 5 5 10 5 5 10 5 Response time average s 2 2 2 2 2 External calibration weight g 5 000 1 0...

Page 13: ...ertencias para un funcionamiento seguro y sin problemas de la balanza el valor de tensi n impreso en el adaptador de CA debe concordar con la tensi n de red local utilizar s lo bater as block de 9 vol...

Page 14: ...tor de vidrio contra las corrientes de aire tapa superior con el borde hacia abajo Balanzas con platillo de pesada redondo 115 mm Montar componentes soporte del platillo de pesada platillo de pesada B...

Page 15: ...x 30 min despu s de la conexi n a la red Reci n entonces la balanza ha alcanzado la temperatura necesaria de funcionamiento Ejemplo Determinar valor de peso Ajustes previos los ajustes de f brica Paso...

Page 16: ...2 1 4 Cantidad piezas de referencia 20 c digo 3 3 3 Paso Pulsar tecla Indicaci n salida 1 en caso dado ON OFF encender bal 2 en caso dado t 0 0 g tarar balanza 3 informaci n indicar FUNCTION ref 20 br...

Page 17: ...tm tico aparece como lectura definitiva en la unidad de peso seleccionada e identificado mediante tri ngulo La cantidad de mediciones puede ser ajustada en el men ver en Ajustes previos La indicaci n...

Page 18: ...idades diferentes Ajustar en el men programa de aplicaci n Conmutaci n de unidades ver en Ajustes previos c digo 2 1 2 Conmutaci n de unidades Item de men Unidad Conversi n S mbolo 1 7 2 o 3 1 2 o Gra...

Page 19: ...s sinopsis o ajuste de f brica ajuste de usuario Men 1 Pesar 1 1 Adaptaci n 1 1 1 ambiente muy de filtro tranquilo 1 1 2 o ambiente tranquilo 1 1 3 ambiente intranquilo 1 1 4 ambiente muy intranquilo...

Page 20: ...pesada no fue observada p ej tare con la tecla t puesta a cero calibre con lectura cero platillo con carga e 09 con bruto cero sin tara Tare la balanza e 10 tecla t bloqueada con reci n despu s de bor...

Page 21: ...CA de la tensi n de red asegure la balanza contra el uso ulterior Un funcionamiento libre de riesgo del adaptador de CA ya no se garantiza si el adaptador de CA presenta da os visibles el adaptador de...

Page 22: ...ibilidad g 0 1 0 1 0 1 0 1 1 Rango de tara substractivo g 6100 3100 1500 610 6 Repetibilidad g 0 1 0 1 0 1 0 1 1 Linealidad g 0 2 0 2 0 2 0 2 2 Deriva de sensibilidad dentro de 10 30 C K 0 5 10 5 1 10...

Page 23: ...98648 012 04 Publication No WXX6050 al03091 W1A000 BOECO BBL GmbH Co scientific equipment Brandstwiete 4 20457 Hamburg t 040 3256270 f 040 32562741 E Mail Boeckel Co t online de Boeckel Co...

Reviews: