STP-603/1003 Series Instruction Manual
1) When securing the inlet port with bolts
1) When securing the inlet port with bolts
Refer to Table 4.2 for torque in pump abnormality and recommended
securing bolts.
Secure the inlet port flange with all of the boltholes of the size specified in
the Inlet Port Flange Standard.
Secure the base with all 8 screw-holes for legs or all 8 attached legs.
Follow "CAUTION" on page 4-9 about legs and bolts for securing the base.
Make sure that the recommended securing bolt may be different
depending on the method of securing the base.
Installation
stallation
Table 4.2 Maximum Torque predicted and Recommended securing bolt for inlet port flange
Model of TMP
STP-603
Type of flange
VG150
ISO160F/ISO160
ICF203
Torque in pump
abnormality [Nm]
1.5×10
4
1.5×10
4
1.5×10
4
Base(8 positions)securing
No Yes No Yes No Yes
Type of bolt
Standard
Standard
Standard Standard Standard Standard
Type of steel
Stainless steel Stainless steel Stainless steel Stainless steel Stainless steel Stainless steel
Recommended
securing bolt
for TMP Flange
Strength
*1
70 or more
70 or more
70 or more
70 or more
70 or more
70 or more
Model of TMP
STP-1003
Type of flange
VG200
ISO200F/ISO250F/ISO200
ICF253
Torque in pump
abnormality [Nm]
1.5×10
4
1.5×10
4
1.5×10
4
Base(8 positions)securing
No Yes No Yes No Yes
Type of bolt
Standard
Standard
Standard Standard Standard Standard
Type of steel
*1
Stainless steel Stainless steel Stainless steel Stainless steel Stainless steel Stainless steel
Recommended
securing bolt
for TMP Flange
Strength
*1
70 or more
70 or more
70 or more
70 or more
70 or more
70 or more
Recommended fitting
bolt for flange
Secure the base
(a) When the base is not secured
(b) When the base is secured
Figure 4.5 Example of securing the STP pump
(When securing the inlet port with bolts)
4-10
*1
Refer to ISO898-1(JISB1051), ISO3506(JISB1054) and AMS6419(Aerospace Material Specification)