24.2 Common Problems in the Device and Corresponding Solutions   

Problem 

Possible Cause 

Solution(s) 

The device does 

not  work  when 

it is turned on 

The Auto On / Off feature is 

enabled   

Take  a  few  deep  breaths  with  the 
mask  on,  and  the  device  will  start 

automatically 

Power  is  not  connected 
properly 

Ensure that the power adapter, and 
the device are connected properly 

There is no voltage 

Check  whether  a  power  outage 
occurs by turning on a light or other 

means.  If  you  are  sure  the  fuse  in 

the  device  is  broken,  contact  your 
equipment supplier for repair 

Cannot find any cause 

Contact your equipment supplier   

The  device  is 

working, but the 
pressure  inside 

the mask differs 

from  the  set 
treatment 

pressure   

The  tube  is  not  connected 
properly 

Reconnect the tube properly 

There  may  be  holes  in  the 
mask  or  pressure  sensing 

tube 

Contact your equipment supplier 

It is a faulty device 

Contact your equipment supplier 

The  effects  of  degraded 

sensors  and  electrodes,  or 
loosened electrodes 

Contact your equipment supplier 

The 

device 

produces  very 

low pressure 

The  air  inlet  of  the  device 
may be blocked 

Replace  the  air  filter  (see  18.4 

Replacing  the  Air  Filter),  and  clean 
the air inlet. Make sure the air inlet is 

unblocked 

The treatment pressure has 

been changed accidentally   

Contact your physician 

When  the  Ramp  feature  is 
enabled, it takes some time 

for  the  initial  pressure  to 

rise  to  the  treatment 
pressure. This is normal 

If  necessary,  disable  the  Ramp 
feature, or set the ramp time shorter 

The device is in 

standby, 

and 

will not start 

The  operating  system  of 

the  device  needs  to  be 
readjusted or restarted 

Unplug the power cord of the device, 

and re-plug it 20 seconds later 

Summary of Contents for M1 Mini

Page 1: ...User Manual Sleep Apnea Therapy Device and Accessories Auto CPAP System M1 Mini 0123...

Page 2: ...Mask 12 11 6 Using Oxygen with the Device 14 11 7 Establish Bluetooth connection 14 11 8 Setting 14 11 9 Starting Treatment 15 12 Routine Use 15 12 1 Connecting the Tubing 15 12 2 Adjusting the Tubing...

Page 3: ...19 Traveling with the Device 22 19 1 Traveling 22 19 2 Traveling by airplane 22 20 Transferring the Device to Another Patient 23 21 Reordering 23 22 Technical Support 23 23 Disposal 23 24 Troubleshoot...

Page 4: ...orized Representative in the European Community AC Power European CE Declaration of Conformity DC Power Bluetooth logo IP22 12 5 mm Diameter Dripping 15 tilted Nonionizing radiation Serial Number of t...

Page 5: ...e as possible To ensure that you receive the safe effective therapy prescribed for you use only BMC accessories WARNINGS The device is intended for adult use only The device is not intended for life s...

Page 6: ...itions before treatment active epistaxis upper gastrointestinal bleeding before treatment coma or impaired consciousness making the use of mask during therapy impossible giant vocal fold polyp etc Rel...

Page 7: ...70 C 13 F to 158 F Humidity 93 Non condensing 93 Non condensing Atmospheric Pressure 760 1060 hPa 760 1060 hPa Mode of Operation Continuous Work Mode CPAP AutoCPAP AC Power Consumption 100 240 V AC 5...

Page 8: ...le the average flow rate at the patient end should be greater than 80 of the corresponding flow value in the table Air Tubing Air tubing Length Inner diameter Tubing 6 ft 1 83 m 15 mm Tubing 6 ft 1 83...

Page 9: ...n air pressure no less than the prescribed one based on the patient s needs 7 Glossary Apnea A condition marked by the cessation of spontaneous breathing AutoCPAP Adjust CPAP pressure automatically to...

Page 10: ...more comfortably Reslex A therapy feature that is enabled by you or your home care provider to provide pressure relief during exhalation Standby State The state of the device when power is applied bu...

Page 11: ...ith the User Manual According to the power adapter standards of different countries different power adapters are configured IMPORTANTS If any of the above parts are missing contact your home care prov...

Page 12: ...Bluetooth is off press this button to turn on the Bluetooth function when Bluetooth is on double click this button to turn off the Bluetooth function The indicator light is blue Air Outlet Deliver pre...

Page 13: ...s entered the enclosure disconnect the power cord and discontinue use Contact your home care provider immediately If the room temperature is warmer than 95 F 35 C the airflow produced by the device ma...

Page 14: ...leading to the malfunctioning of the device Air must flow freely around the device for it to work properly 11 3 Installing the Air Filter and Filter Cap 1 Attach the air filter to the filter cap as s...

Page 15: ...ifficult to disconnect the power supply Do not stack too long cables or tubing at the head of the bed which may entangle the head or neck of the patient during sleeping CAUTION Inspect the cord of pow...

Page 16: ...Position the exhalation port so that the vented air is blowing away from your face Connect the mask s connector to the exhalation port If you are using a full face mask a mask covering both your mouth...

Page 17: ...than 1 m from the device When using oxygen with this system a Pressure Valve must be placed in line with the patient circuit between the device and the oxygen source The pressure valve helps prevent t...

Page 18: ...manual for the mask CAUTION Before each use examine the tubing for any damage or debris If necessary clean the tube to remove the debris Replace any damaged tube Make sure that the mask does not leak...

Page 19: ...ng which the initial pressure rises to the prescribed treatment pressure can be adjusted The numbering increases or decreases by five minutes The LightTrip App displays a real time countdown of the re...

Page 20: ...720 days Set the mask remind cycle it will remind the user once the cycle is over The setting increment is 30 days Tubing Off 30 days 720 days Set the tubing remind cycle it will remind the user once...

Page 21: ...vice is connected with LightTrip App by Bluetooth a prompt of Please change the air filter will appear in LightTrip App Please replace the tubing When the remind cycle of tubing is set the status ligh...

Page 22: ...ll score of 10 Compliance 0 10 10 The compliance of a single day is calculated according to the proportion of effective days used in the past week while the compliance of multiple days is calculated a...

Page 23: ...servicing should only be performed by an authorized service agent In order to prevent cross infection of patients or contamination of equipment BSF Breathing System Filter that meet ISO 23328 1 2003 a...

Page 24: ...device with a soft slightly damp cloth CAUTION The device can only be used after the enclosure is dry so that no moisture enters the device 18 3 Cleaning the Tubing 1 Remove the tubing from the devic...

Page 25: ...ments filled and signed by your physician about the device If you plan to travel by air remember to bring the multi language emergency documents about respiratory therapy in case that the border and c...

Page 26: ...al or harsh sounds if it has been dropped or mishandled if the enclosure is broken or if water has entered the enclosure discontinue use Contact your home care provider If the device malfunctions cont...

Page 27: ...k size or model may not be correct or the mask is not positioned correctly thereby leading to air leakage Narrow the distance between the forehead support of the mask and the forehead Note that adjust...

Page 28: ...cur Probably because the patient sleeps with his or her mouth open and the pressurized air goes out via the mouth leading to blockage in the respiratory tract Use a chin strap to prevent the mouth fro...

Page 29: ...nected properly Reconnect the tube properly There may be holes in the mask or pressure sensing tube Contact your equipment supplier It is a faulty device Contact your equipment supplier The effects of...

Page 30: ...ded for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance...

Page 31: ...ituations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cor...

Page 32: ...ted maximum output of transmitter W 150 kHz 80 MHz p d 17 1 80 MHz 800 MHz p d 35 0 800 MHz 2 5 GHz p d 70 0 0 01 0 12 0 04 0 07 0 1 0 37 0 12 0 23 1 1 17 0 35 0 70 10 3 70 1 11 2 22 100 11 7 3 50 7 0...

Page 33: ...etween RF wireless communications equipment and the device as recommended below according to the maximum output power of the communications equipment Frequency MHz Maximum Power W Distance IEC 60601 T...

Page 34: ...anufacturer of the equipment or system as replacement parts for internal components may result in increased emissions or decreased immunity of the equipment or system The device may be interfered with...

Page 35: ...D DISCLAIMS ALL LIABILITY FOR ECONOMIC LOSS LOSS OF PROFITS OVERHEAD OR CONSEQUENTIAL DAMAGES WHICH MAY BE CLAIMED TO ARISE FROM ANY SALE OR USE OF THIS PRODUCT SOME STATES DO NOT ALLOW THE EXCLUSION...

Page 36: ...391263...

Reviews: