11.6 Using Oxygen with the Device
Oxygen may be added at the mask connection. Please observe the instructions listed below
when using oxygen with the device.
WARNINGS!
• Connect the oxygen tube to the oxygen inlet of the mask.
• The oxygen supply must comply with the local regulations for medical oxygen.
• Turn on the device before turning on the oxygen. Turn off the oxygen before turning off the
device. Explanation of Warning: When the device is turned off, but the oxygen flow still exists,
oxygen may accumulate within the device's enclosure and pose a fire hazard. Turning off the
oxygen before turning off the device will prevent oxygen accumulation in the device and
reduce the risk of fire. This warning applies to most CPAP devices.
• Oxygen supports combustion. Keep the device and the oxygen container away from heat,
open flames, any oily substances, or other sources of ignition. DO NOT smoke in the area near
M1 Mini or the oxygen container.
• Sources of oxygen should be located more than 1 m from the device.
• When using oxygen with this system, a Pressure Valve must be placed in-line with the patient
circuit between the device and the oxygen source. The pressure valve helps prevent the
backflow of oxygen from the patient circuit into the device when the unit is off. Failure to use
the pressure valve could result in a fire hazard.
• Do not connect the device to an unregulated or high pressure oxygen source. The pressure of
oxygen source does not exceed the work pressure of the device.
11.7 Establish Bluetooth connection
Connect the device to a power supply. Press the Bluetooth button on the main device and the
Bluetooth indicator light flashes. Then open the LightTrip App, click the Bluetooth icon, and
start to search the device, select the device (name is device serial number, please check the
device nameplate information) in the device list for Bluetooth connection. If the Bluetooth is
connected successfully, the Bluetooth working status light of the main device will be kept on.
11.8 Setting
Click "Settings" and "Accessories" in the LightTrip App to enter the corresponding Settings
interface respectively. See "13. Navigating the Patient Menu" in this manual for detailed of the
interface.