M1 Mini User Manual
25 / 32
Problem
Possible Cause
Solution (s)
Discomfort
due
to
inability
to
adapt to the
treatment
pressure
The
patient
will
feel
uncomfortable
when
the
treatment pressure is higher than
13 hPa. However, the treatment
pressure is determined according
to the patient's conditions, and
cannot treat sleep apnea if the
treatment pressure is set too low
It takes a maximum of four weeks
to adapt to pressurized air.
Relax and breathe through the
nose. If the problem still exists,
contact your physician
Obstructive
sleep apnea
symptoms
recur
Probably because the patient
sleeps with his or her mouth
open, and the pressurized air
goes out via the mouth, leading
to blockage in the respiratory
tract
Use a chin strap to prevent the
mouth from opening during sleep,
or use a full-face mask. Contact
your physician for details
The device is
too noisy
The tube is not connected
properly
Reconnect the tube properly
Air delivered
from
the
device
is
abnormally
hot
The air inlet of the device may be
partially blocked, leading to
insufficient airflow into the device
Replace the air filter (see 18.4
Replacing the Air Filter), and
clean the air inlet
Place the device in an area where
air flows freely, and make sure
the device is at least 20
centimeters away from the wall,
curtain, or other things