G2S BPAP Series User Manual V2.1
40 / 43
23. EMC Requirements
The cables must be provided by BMC. The information of each cable is as follows:
(1) Power Cord: 1800 mm ± 30 mm, unshielded.
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The
user of the device should ensure that it is used in such an environment.
Emissions Test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3
Complies
WARNINGS!
• This device may be interfered with by other equipment, even if that other equipment
complies with CISPR EMISSION requirements.
• Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the G2S
BPAP series, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
• During operation of the device, due to electrostatic interference, the following phenomena
may occur: (1) Temporary loss of function or degradation of performance, such as abnormal
screen display, etc. The device will recover to normal after being restarted; (2) Automatic
restart of the device. These phenomena will not affect the normal use of the device, and will
not cause permanent performance degradation or function loss of the device.