G2S BPAP Series User Manual V2.1
30 / 43
High Pressure!!!
High
Priority
Alarm
When the airflow is on, an audible alert
will sound if the airway pressure
exceeds the warning threshold; the
screen will display “
High Pressure!!!
”.
Note:
The thresholds for different
models:
Off, 5
~
21 hPa applies to G2S B20T,
in 0.5 hPa increments, the default
setting is “
20 hPa
”.
Off, 5
~
26 hPa applies to G2S B25T,
G2S B25VT and G2S LAB, in 0.5 hPa
increments, the default setting is “
25
hPa
”.
Off, 5
~
31 hPa applies to G2S B30T,
G2S B30VT and G2S B30AT, in 0.5 hPa
increments, the default setting is “
30
hPa
”.
Low Pressure!!
Middle
Priority
Alarm
When the airflow is on, an audible alert
will sound if the airway pressure is
below the warning limen; the screen
will display “
Low Pressure!!
”.
Note:
The limens for different models:
Off, 3
~
19 hPa applies to G2S B20T,
in 0.5 hPa increments, the default
setting is “
4 hPa
”.
Off, 3
~
24 hPa applies to G2S B25T,
G2S B25VT and G2S LAB, in 0.5 hPa
increments, the default setting is “
4
hPa
”.
Off, 3
~
29 hPa applies to G2S B30T,
G2S B30VT and G2S B30AT, in 0.5 hPa
increments, the default setting is “
4
hPa
”.
Low RR!!!
(only applies to G2S
B20T, G2S B25T,
G2S B25VT, G2S
B30T, G2S B30VT,
G2S
B30AT
and
G2S LAB)
High
Priority
Alarm
When the airflow is on, an audible alert
will sound if the respiratory rate is
below the limen; the screen will display
“
Low RR!!!
”.
Setting range: Off, 4
~
40 BPM, in 1
BPM increments, the default setting is
“
6 BPM
”.
Note:
This function is available under
the work mode of S/T or T.
Leak!!
Middle
Priority
Alarm
When the airflow is on, an audible alert
will sound if the air leak rate exceeds
150 L/min; the screen will display
“
Leak!!
”.
The alarming duration time is no less
than 30 s.