G2S BPAP Series User Manual V2.1
31 / 43
Low SpO
2
!!!
High
Priority
Alarm
When SpO
2
Kit is applied, an audible
alert will sound when the value of SpO
2
is lower than the warning threshold; the
screen will display “
Low SpO
2
!!!
”.
Setting range: Off, 70%
~
100%, in
1% increments, the default setting is
“
85%
”.
Note:
This function is available only
when the device is equipped with SpO
2
Kit.
Mask Blocked!!
(only applies to G2S
B20T, G2S B25T,
G2S B25VT, G2S
B30T, G2S B30VT,
G2S B30AT and G2S
LAB)
Middle
Priority
Alarm
When the airflow is on, an audible alert
will sound if the vents of the mask are
blocked; the screen will display “
Mask
Blocked!!
”.
Low MV!!
(only applies to G2S
B20T, G2S B25T,
G2S B25VT, G2S
B30T, G2S B30VT,
G2S B30AT and G2S
LAB)
Middle
Priority
Alarm
When the airflow is on, an audible alert
will sound if the minute volume is below
the warning limen; the screen will
display “
Low MV!!
”.
Setting range: Off, 1
~
30 L/min, in 1
L/min increments, the default setting is
“
1 L/min
”.
Low Input Voltage!!
Middle
Priority
Alarm
If the voltage supplied by power adaptor
is lower than 22 V, an audible alert will
sound and the screen will display “
Low
Input Voltage!!
”.
High RR!!
(only applies to G2S
B20T, G2S B25T,
G2S B25VT, G2S
B30T, G2S B30VT,
G2S B30AT and G2S
LAB)
Middle
Priority
Alarm
When the airflow is on, an audible alert
will sound if the respiratory rate exceeds
the threshold; the screen will display
“
High RR!!
”.
Setting range: Off, the setting value of
Low RR
~
80 BPM, in 1 BPM
increments, the default setting is “
40
BPM
”.
Note:
This function is avaliable under
the work mode of S/T or T.
Humidifier Failure!!
Middle
Priority
Alarm
When humidifier is applied, an audible
alert will sound when the humidifier fails
to work in 10 minutes; the screen will
display “
Humidifier Failure!!
”.
Please Change
Filter!
Low
Priority
Alarm
When the Filter Alert feature is enabled,
an audible alert will sound if the preset
replacement time reaches but without
replacing the air filter; the screen will
display “
Please Change Filter!
”. The
default setting is “
Off
”.