E-20A-H-O / E-20AJ-H-O User Manual V2.1
4 / 36
Eng
lis
h
5. Specifications
Device Size
Dimensions: 170 mm × 180 mm × 118 mm, or 290 mm × 180 mm × 134 mm (with the
humidifier)
Weight:1.5 kg, or 2.5 kg (with the humidifier)
Product Use, Transport and Storage
Operation
Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing
15% to 93% Non-condensing
Atmospheric Pressure: 760
~
1060 hPa
760
~
1060 hPa
Mode of Operation
Continuous
Work Mode
CPAP, Auto
SD Card
The SD card can record patient data and fault information.
AC Power Consumption
100
-
240 V AC, 50 / 60 Hz, 2.0 A max
Main Device offer to USB Communications Port
5 V
2.0 A
Main Device offer to Humidifier
24 V
1.5 A
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
4 to 20 hPa (in 0.5 hPa increments), ≤30 hPa under single fault conditions.
Pressure Display Accuracy
± (0.5 hPa
+
4%)
Static Pressure Stability
± 0.5 hPa
Ramp
The ramp time ranges from 0 to 60 minutes.
Summary of Contents for E-20A-H-O
Page 1: ......
Page 2: ...English User Manual RESmart GII Auto CPAP System E 20A H O E 20AJ H O 0123...
Page 41: ...Espa ol Manual del Usuario RESmart GII Auto CPAP Sistema E 20A H O E 20AJ H O 0123...
Page 80: ...Italiano Manuale Utente RESmart GII Auto CPAP Sistema E 20A H O E 20AJ H O 0123...
Page 118: ...390400...