Troubleshooting
If the Bluezone Logo Status Panel is not
illuminated, then the PK-220 is not
being powered. Check the following list
for possible causes and solutions:
1. Check that unit is powered.
a.
Power “Off” unit by pressing
the power switch located on
the control panel.
b. Plug unit in.
c. Verify the outlet has power
(if outlet is controlled by a
switch, check that circuit
switched on).
d.
Turn unit “On.”
2. Make sure the cover is on and
the cover latch engages the
cover switch.
See Figure 16 for proper closure of the
cover.
3.
Check whether the u
nit’s circuit
breaker has tripped.
a. Visually check the circuit
breaker on the side of the
unit. If the breaker has
popped out, follow b, c, d
below:
b.
Power “Off” unit using the
power switch located on the
power entry module.
c. Reset circuit breaker by
pushing the visible indicator
back into the circuit breaker.
d.
Turn unit “On” using the
power switch located on the
power entry module.
Danger
DO NOT
override the cover switch when the cover is removed. UV light will be emitted.
UV radiation is dangerous for both the eyes and skin. Do not operate the UV-C emitter
outside the appliance.
Summary of Contents for PK-220
Page 4: ...Document BPI PM013 00 3 Key Components Figure 1 Bluezone PK 220 Key Components ...
Page 11: ...Document BPI PM013 00 10 Figure 8 PK 220 Floor Mounted Figure 9 PK 220 Wall Mounted ...
Page 19: ...Document BPI PM013 00 18 Figure 15 Display Sequence in Time ...
Page 26: ...Document BPI PM013 00 25 Figure 17 Remove Service Cover Figure 18 Remove Bulb Clip Screws ...
Page 28: ...Document BPI PM013 00 27 Figure 21 Slide Bulb with Bulb Clip into the Bulb Support Bracket ...
Page 29: ...Document BPI PM013 00 28 Figure 22 Install Bulbs into Sockets Figure 23 Install Service Cover ...