BVC6100LM Instruction Manual v.01
- 22 -
「Polarity」: Select polarity of external trigger signal
Active Low : Exposure is started at falling down edge of external trigger
in Shutter Select Mode
Exposure time is set by the LOW level period of extrnal trigger
in PWC Mode
Active High : Exposure is started at rising up edge of external trigger
in Shutter Select Mode
Exposure time is set by the HIGH level period of extrnal trigger
in PWC Mode
6.1.2 Line rate
Line rate setting is available in Internal sync.
Input Line rate at 「Line Rate」 window the click「SET」 or Enter.
「Line Period」 window is displayed Line period cycle of Line rate set the above.
「Line Period」 window is unable to input, just displaying only.
Line rate setting range 10Hz~20000Hz
です。
It is valuable to input by 1Hz step.
6.1.3 Exposure time
It is available to set exposure time in Shutter Select Mode for Trigger mode.
Exposure time setting range:
「Source」が External Trigger OFF:10(us)~「Line Period」(us) – 9.52(us)
「Source」が External Trigger ON :10(us)~External trigger cycle – 9.52(us)
It is valuable to input by 0.01us.
※Exposure time shuould be set within LinePeriod(us), in case of exposure time setting over
the LinePeriod time, it is necessary to decrease LineRate(Hz) by appropriate level.