background image

2.0    Cleaning

Cleaning the conductivity pen probe periodically ensures accurate readings. 

The probe is cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif” 

a trade name for a liquid scourer cream used in home bathrooms and kitchens. 

Similar products are called “Liquid Vim”, “Soft Scrub”, “Cif cream”, or “Viss”. 

Never use scented varieties as they contain oils that contaminate the probe. 

Follow the steps below to clean the probe. 

Remove shroud.

Hold the body and pull the shroud off. Holding your hand 

around the shroud for a few seconds will help with removal.

Clean probe face. 

Place one or two drops of probe cleaner onto the probe 

face and rub with your finger or Bluelab Chamois firmly and 

vigorously, to clean the probe face. 

If a heavy build up occurs around the temperature sensor 

clean with a soft toothbrush to remove contamination.

Rinse probe.  

Rinse off all traces of cleaner under running water using the 

same finger or other side of Bluelab Chamois.

Check that the water forms a film on the probe face with 

no “beads” of water. If beading is present repeat the 

cleaning process.

Replace shroud and test in a known solution to ensure 

the unit has been adequately cleaned.

Calibration of the probe is required after every clean. 

See the pH calibration in section 6.0. Place storage cap 

back onto the probe.

3.0    Battery replacement

The conductivity pen is powered with 1 x AAA alkaline battery. Do not use 

rechargeable batteries. A low battery warning is indicated by a battery symbol 

appearing on the screen. Only remove the battery cap when the batteries 

require changing. Battery life is expected to be 350 hours.

To remove old battery

Undo battery cap fasteners. Remove battery cap and tip out the 

old battery.

Check for corrosion

Flat batteries may leak and cause corrosion. Check battery 

contacts and the battery for any sign of corrosion. Battery 

contacts should be cleaned first if corrosion is found before 

proceeding to step 3.

Fit new battery

Insert the new batteries positive (+) end down into the body.

Ensure waterproof battery cap seal is clean

Seal will fail if any dirt is present.

Replace battery cap

Tighten fasteners on battery cap until there is no space left 

between the cap and body. 

This ensures the unit remains 

100% waterproof.

Waterproof seal

English

4

1

2

3

4

1

2

3

4

5

Pull

Summary of Contents for conductivity pen

Page 1: ... Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about the scales available on the Bluelab Conductivity Pen 7 Blulelab Probe Care Kits 7 Bluelab limited warranty 8 Contact details 9 Care and use guide a handy solution made easy bluelab com ...

Page 2: ...g Long press to turn off Calibrate button See calibration section Units button Hold until units flash then short press to change units Screen will change back when no buttons have been pressed for 3 seconds Quick guide Low battery warning Appears when batteries are low Check mark tick to indicate successful calibration ATTENTION The instrument is only as accurate as the probe is clean Shroud The c...

Page 3: ...he power button again To change units Hold down the units button for 3 seconds until the conductivity and temperature units start flashing Short press units button again to cycle between unit combinations To exit this mode don t press anything for 3 seconds NOTE You can change units while in hold mode by holding down the units button Rinse conductivity probe To reduce the build up of nutrient salt...

Page 4: ...is present repeat the cleaning process Replace shroud and test in a known solution to ensure the unit has been adequately cleaned Calibration of the probe is required after every clean See the pH calibration in section 6 0 Place storage cap back onto the probe 3 0 Battery replacement The conductivity pen is powered with 1 x AAA alkaline battery Do not use rechargeable batteries A low battery warni...

Page 5: ...played If Err is displayed check the probe is clean and that the calibration solution is fresh and uncontaminated A check mark will appear on the screen to indicate that the calibration was successful The check mark will disappear after 30 days To reset back to factory default remove replace battery NOTE If you need to test or calibrate in a 1500 ppm solution you MUST set the pen to EC then multip...

Page 6: ...ion Check the table in section 4 0 for what solution to use for your selected conductivity unit Screen does not turn on Replace unit 7 0 Technical specifications Measurement Range 0 0 10 0 EC 0 5000 ppm 500ppm TDS 0 7000 ppm 700 ppm 0 50 C 32 122 F Resolution 0 1 EC 10 ppm 700 10 ppm 500 1 C 1 F Accuracy at 25 C 77 F 0 1 EC 50 ppm 500 ppm 70 ppm 700 ppm 1 C 2 F Temperature compensation Automatic O...

Page 7: ...solution Individual nutrient ions have different electrical effects The true ppm of a solution can only be determined by a chemical analysis ppm cannot be accurately measured by an EC meter They are present on Bluelab products as a conversion guide only The conversion is as follows 2 4 EC x 500 1200 ppm 500 scale or 1200 ppm 500 2 4 EC 2 4 EC x 700 1680 ppm 700 scale or 1680 ppm 700 2 4 EC If you ...

Page 8: ...oduct Product repaired or altered by parties other than Bluelab or its authorised agents Product with defaced missing or illegible serial numbers Products not purchased from Bluelab or a Bluelab authorised distributor or reseller How Do You Get Service To begin a warranty claim you must return the Product to the point of purchase with valid proof of purchase as defined above In California you can ...

Page 9: ...7 578 0847 Email support bluelab com get online Looking for specifications or technical advice Visit us online at bluelab com or facebook com getbluelab To watch instruction videos visit our online video library vimeo com bluelab post Bluelab Corporation Limited 8 Whiore Avenue Tauriko Business Estate Tauranga 3110 New Zealand Instruction Manual English PENCON_V01_210916 Copyright 2015 all rights ...

Reviews: