Blue-White APNTU Quick Start Manual Download Page 7

2

Engineering Specifications

Item

APNTU

Turbidity Wavelength

Warm White

Light Source

LED

Turbidity Dual Range

0.001 – 10 / 10 - 40.00 NTU

Turbidity Accuracy

+/- 0.005 NTU or 2% <10NTU

Turbidity Repeatability

+/- 0.001 NTU or 0.5% <10NTU

Measurement Accuracy

0.001 NTU or ± 1% Full Scale

Minimum Resolution

0.001 NTU

Response time

4s after immersion - Turbidity

Compliance

EPA-180.1

Measurement Interval

Continuous Measurement

Display

7-inch LCD Color Industrial Capacitive Touch Screen

Storage Capacity

Built-In 4GB of Ram for Storing up to 1-Million Data/Event Records

Power Requirement

96-260VAC / 50-60 Hz; 10A Fuse; 200 W

Output

2 x 4-20 mA / RS-485 Modbus - RTU / Modbus TCP

Input

2 x 4-20 mA / RS-485 Modbus - RTU

USB

1 x USB host, for data downloading and screen upgrade

Internet

RJ-45 socket, Modbus-TCP

Panel Operational Temperature

40 – 113

o

F (4-45 

o

C)

Storage Temperature

Instrument: -4 – 131

o

F (-20 – 55

o

C) / Sensors 32 – 122

o

F (0 – 50

o

C)

Sample Water Temperature

40 – 104

o

F (4-40

o

C)

Sample Water Pressure

7.25 – 30 psi (0.05 – 0.2MPa)

Installation

WR10 Self-Regulating Flow Reservoir w/Rotameter & Isolation Valve - Included

WR10 Minimum Flow Rate

3.1 g/h (200 mL/minute)

WR10 Maximum Flow Rate

6.2 g/h (400 mL/minute)

WR10 Sample Inlet

1/4 - inch OD

WR10 Sample Outlet

3/4 - inch ID - To Drain 

WR10 Drain

1/2 - inch NPT

Rating

IP-65 Panel-Display / IP-67 Sensors

Regulation

CE / RoHS

Relative Humidity

20% - 90% (No Condensation)

Altitude

<6,561 feet (<2,000 Meter)

Dimensions (HxWxD)

Panel  36.00 H x 20.00 W x 9.42 D inches

Approximate Product Weight

33 lbs

Shipping Dimensions

42” x 26” x 16” (1067H x 661W x 407D mm)

Engineering Specifications

Summary of Contents for APNTU

Page 1: ...Quick Start Guide APNTU Turbidity Analyzer Panel...

Page 2: ...l time measurement display and data logging of Ultra Low Turbidity utilizing proprietary smart sensor technology coupled with a touch screen display and data logging terminal Reference the below dimen...

Page 3: ...trical Schematics Quick Start Guide Tools Needed Follow these instructions before installing your analyzer to avoid failure Read this guide and Instruction Manual before using CAUTION The APNTU is des...

Page 4: ...g to the outlet drain This is the sample water outlet flow This line must be diverted to drain Wall Mount Space The APNTU analyzer panel size is roughly 20 H x 36 W x 10 D in dimension Please accommod...

Page 5: ...e and clean thoroughly Located strut and mount with appropriate tools Check to make sure strut is secure and level 2 4 Connect Electrical and Tubing Use provided electrical diagram to connect power an...

Page 6: ...y necessary when changing parameter settings or performing calibrations To access this now press the icon and refer to Instruction Manual for more instructions regarding login and user management Real...

Page 7: ...U USB 1 x USB host for data downloading and screen upgrade Internet RJ 45 socket Modbus TCP Panel Operational Temperature 40 113o F 4 45 o C Storage Temperature Instrument 4 131o F 20 55o C Sensors 32...

Page 8: ...ssive heat or pressure or allowing the unit to support the weight of related piping BLUE WHITE does not assume responsibility for any loss damage or expense directly or indirectly related to or arisin...

Reviews: