Appendix B : Tip Ejector Height Adjustment
1. Remove the lower part of pipette as described in Section 11.1.
2. Remove the flat-head screw and bushing as shown in Figure 14.
3. Use flat-head screwdriver or other tool to rotate the tip ejector
height adjustment nut. By rotating the nut in counter-clockwise
direction, the tip ejector will move closer to the mouth of tip fitting.
This height adjustment will enable the pipette to accommodate to
a wider range of tips.
Tip ejector height
adjustment nut
Flat-head screw
Figure 14 Tip ejector height adjustments