6. SYS (System) Mode
SYS mode is used to set up the function settings before pipetting. Once
users select and store these settings, they will no longer need to set up
the settings before every pipetting. There are 7 function settings
(Aspirating Speed, Dispensing Speed, Acoustic Signals, Tip Ejection
Alarm, Blow-out, Reverse Pipetting and Program Recall) in SYS mode.
Note:
1. These 7 function settings are sequential procedures. To store
these settings in memory, you have to complete the whole setting
procedure and press «SET» when the LCD displays
.
2. You can use the
«Up/Down» button to move back and forth to
the previous function settings.
6.1. Aspirating Speed
There are 5 speeds available for both aspirating and dispensing:
1. Press
«EXIT» to enter the function mode selection. Then
press
«Up/Down» repeatedly until the
symbol starts
flashing.
2. Press
«SET» to confirm the selection of
mode and the
aspirating speed symbol
will start flashing.
3. Press
«Up/Down» to change the aspirating speed ( :
Slowest, : Fastest).
4. Press
«SET» to confirm the speed selection and move to the
“Dispensing Speed” function setting.
13