3.
Sequential Aspirating (SA) function will not perform Reverse
Pipetting even it is selected in System (SYS) mode.
10.2. SD (Sequential Dispensing) Mode
The BluePette performs repetitive dispenses of the selected volumes.
When the first aspiration is completed, the pipette will sequentially
dispense the aspiration volume.
1. Press
«EXIT» to enter function mode selection.
2. Press
«Up/Down» repeatedly until
symbol starts
flashing.
3. Press
«SET» to confirm the selection.
4. Press
«Up/Down» repeatedly until
symbol starts flashing.
5. Press
«SET» to confirm the selection.
6. Under
symbol, “01” will appear. Press
«Up/Down» to
set the volume of the first dispensing.
7. Press
«SET» to confirm the selection.
8. Repeat steps 6 and 7 to set the sequential dispensing volumes.
The total dispensing volume cannot exceed maximum volume
range.
9. Press
«SET» to confirm the selection.
10. The count “01” and the total aspirating volume will appear on the
display. The pipetting direction and speed
will flash to indicate
the pipette is ready for liquid pick-up. Press the
START
button to
aspirate the samples. When the aspirations are completed, the
dispensing count and volume will appear and
will flash to
indicate that the pipette is ready for dispensing.