127
CHAPTER 10: Calibration
10.5 Resistance Calibration
Resistance calibration should be performed on a periodic basis in order to
maintain maximum accuracy. Resistance calibration is most sensitive to
temperature changes. Allow the main unit to stabilize to the operating
environment before performing the calibration procedue. Calibration
should be performed whenever a different Performance Module is
installed.
To calibrate Resistance:
1. Connect the SCAN-LAN Vt to the Remote Unit.
2. Set the dial to CAL.
3. Highlight Resistance in the Calibrate menu by using the arrow keys.
4. Press CALIBRATE (F4) or ENTER. The Calibrate Resistance screen
will appear. The bar graph across the bottom of the display will show
the progress of the calibration.
5. When finished, the screen returns to the CALIBRATE menu.
10.6 Main-Unit Calibration
This calibration mode is intended for factory use only.