CHAPTER 10: Calibration
for remote operation. However, if Remote Calibration is performed, it’s
important to verify that the calibration was completed properly.
To calibrate the remote unit:
1. Connect the main unit to the CAT 3 calibration coupler, using the
Modular Plug Performance Module. (If you are using the Modular Jack
Performance Module, use the four-foot CAT 5 patch cord.)
2. Connect the remote unit to the open end of the CAT 3 calibration
coupler using the Modular Plug Performance Module.
3. Set the remote unit’s dial to REMOTE.
4. Set the main unit’s dial to CAL.
5. Highlight REMOTE in the Calibrate menu.
6. When the test is finished, the screen returns to the Calibrate menu.
NOTE
In order to ensure correct calibration, be sure to switch the roles of both
units, and repeat the procedure above.
To verify the calibration:
1. Connect the main unit to the CAT 3 calibration coupler, using the
Modular Plug Performance Module.
2. Connect the remote unit to the open end of the CAT 3 calibration
coupler using the Modular Plug Performance Module.
3. Turn the knob to Quick
√
. Press the up or down arrow key to select the
CAT 5 Basic Link test standard. Press ENTER or SAVE (F4) to start the
test.
4. Once the test is complete, press VIEW (F3). Press the up or down arrow
key until the Attenuation results are displayed. The Attenuation results
should display values greater than 0.0 dB. If 0.0 dB is displayed for any