Battery Capacity Analyzer User Manual
37
Press OPTION to TRG
Arrow Keys to Set
I Min = 00323 mA
I Med = 07869 mA
6.4 Load Calibration
This is a similar process to Current calibration with only 3 calibration points. The button is used to trigger the 603B
placing a load on the power supply. Monitor the maximum current used as displayed on the power supply current meter.
Write the maximum value down. Enter this 4 digit value in the row. Repeat until all three calibration points are adjusted.
Note that the value in the 603B is in mA.
From the 3 part calibration screen move the point to the row “Load Calibration”.
25. Press the button to select Load calibration
26. When ready, press the button to trigger the 603B.
27. Use the button to move the cursor into the Load field and enter the max current displayed on the power supply.
28. Repeat these steps until all Min, Med, Max current values have been calibrated. Note this is spread across
2 separate screens.
29. Complete a fan test and verify the 603B fan runs smoothly.
30. Create a calibration label and apply to the 603B.
The Calibration procedure is complete.
I Med 4 = 01462 mA
I Max 4 = 02896 mA
Current Calibration 4 of 4 screens
Voltage Calibration
Current Calibration
Load Calibration
Load Calibration 1 of 2
I Max = 11350 mA
Load Calibration 2 of 2