11
1
1.3 Specifications
Model
4007B
4013B
Frequency Characteristics
Sine
0.1 Hz
– 7
MHz
0.1 Hz
– 12
MHz
Square
0.1 Hz
– 7
MHz
0.1 Hz
– 12
MHz
Triangle
0.1 Hz
– 1 MHz
Accuracy
0.01% ± 0.186 Hz
Resolution
5 digits or 100 mHz
Output Characteristics
Amplitude Range
10 mV to 10 Vp-p (into 50
Ω)
20 mV to 20 Vp-p (open
circuit)
Resolution
3 digits (1000 counts)
Amplitude Accuracy ± 2 % ± 20 mV of
programmed output from 1.01
V
– 10 V
Flatness
± 1 dB to 12 MHz
Offset Range
± 4.5 V (into 50 Ω)*
Offset Resolution
10 mV, 3 digits
Offset Accuracy
± 2 % ± 10 mV (into 50 Ω)
Output Impedance
5
0 Ω
Output Protection
Protected against short circuit
or accidental voltage
practically available in
electronic laboratories,
applied to the main output
connector
Waveform Characteristics
Harmonic Distortion DC
– 100
DC
– 100
Summary of Contents for 4007B
Page 1: ...Model 4007B 4013B DDS Function Generator USER MANUAL...
Page 34: ...34...
Page 35: ...35...