2052 User Guide
(BB1804-C)
15
Make sure that the marker on the water standoff collar is aligned with the marker
on the transducer.
4
Pull a water standoff condom over the transducer, attaching it to the groove on
the water standoff collar. Check the water standoff condom for tears.
5
Secure the water standoff condom in place with 2 or more silicone rubber bands
in each groove (silicone rubber bands are supplied with water standoff condoms
UA0037).
6
Fill the syringe with degassed water. Use the 2-way stopcock and plastic
extension tube to attach the syringe to the Luer lock valve on the water standoff
collar.
7
Fill the water standoff condom with approximately 50ml of degassed water.
align the two markers
water standoff
collar
two markers aligned
2-3 silicone rubber bands in
each of the two grooves
wire clip
Luer lock valve