24
January 2014
2052 User Guide
(BB1804-C)
Figure 15. Removing the outer cover.
3
Remove the tape from the inner cover.
Figure 16. Removing the tape.
4
Remove the inner cover by pulling it inside out up over the transducer tip.
Figure 17. Removing the inner cover.
Clean Immediately after Use
If biological materials are allowed to dry on the transducer or attachments,
disinfection and sterilization processes may not be effective. Therefore, you must
clean attachments and transducers immediately after use.
Use a suitable brush to make sure that biological material and gel are removed from
all channels and grooves, when using reusable water standoff system (including the
collar and wire clip attachment). See
Care, Cleaning & Safety
for cleaning and
disinfection instructions.