16
May 2019
20R3 User Guide
(16-01278-EN-07)
Figure 8. Putting on the inner cover.
4
Create a tape handle: Partly remove the protective strip from the tape so only the
first cm of the tape is exposed. Then bend the first (sticky) cm in half and tape
the sides together to create a 0.5 cm “handle”. (See Fig. 9.)
Figure 9. Creating a tape “handle”.
NOTE:
Make sure you do not tape more than the first cm together. The tape will be
too short to fit around the transducer.
5
Push the non-handle end of the tape through the opening in the inner cover and
wind the tape tightly around the cover and then around the water tube that forms
the opening.
Make sure that when you wind the tape around, it covers the top of the water
tube on the inner cover by 1-2 mm.
Follow the steps shown in Fig. 10. The tape must be tight so that when you put
water in the standoff, it doesn’t run out under the tape.
Inner cover on correctly
Expose 1 cm of the tape.
Bend tape to
make little
handle.
0.5 cm
handle