20
May 2019
20R3 User Guide
(16-01278-EN-07)
Inserting the transducer and reusable water standoff system
Reusable
water
standoff
system
Make sure that the entire outer surface of the water standoff cover is thoroughly
lubricated with a glycerine-based lubricant.
Insert the transducer just until the tip of the transducer is at the far end of the
rectosigmoidoscope.
The end of the rectosigmoidoscope nearest you will be at the 200 mm mark on the
transducer if you are using either the UA0672 or UA0673 rectosigmoidoscope. If
you are using the single-use rectosigmoidoscope UA0683 then it will be at the
190 mm mark.
Hold the transducer in position and pull the rectosigmoidoscope slowly back toward
yourself until the rim touches the water standoff collar UA0671.
NOTE:
It is important to hold collar and rectoscope tight together.
Click the wire clip over the rectosigmoidoscope to hold it in place with the water
standoff collar. Do not reposition the transducer relative to the patient while you are
doing this.
rectosigmoidoscope
wire clip
click the wire clip
over the
rectosigmoidoscope