8.2 Over Temperature Safety Cut-out
The unit is fitted with a non-resettable thermal fuse which blows if the
heater plate temperature exceeds 235 ºC.
In the unlikely event of a fault condition, this fuse will permanently disable
the heater to protect the user and unit from injury and damage. If this
occurs, the heater plate will remain cold and the
xxsealer
will need to be
returned for repair.
8.3 Routine Cleaning and Inspection
The unit casework should be cleaned and inspected at regulator internals,
and whenever contamination or spillage occurs, as follows:
1. Switch off the unit and disconnect the power before performing
any inspection checks or cleaning.
2. Before cleaning, always inspect the unit casework, heater plate
and moving parts for any signs of wear, damage, cracks or other
defects.
3. Clean the casework using a damp cloth soaked with a
disinfectant solution (such as Virkon), whilst wearing suitable
PPE.
4. Carefully clean the heater plate surface to remove any debris or
sealing material.
5. Remove any debris or fluff from around or between the moving
parts of the handle mechanism and sliding plate carrier.
6. Clean the display and buttons, taking care to avoid over wetting.
7. Check that the ventilation holes on the rear of the unit are clear
of dust and fluff build-up.
8. Check and clean the adaptor blocks.
DANGER
After cleaning, ensure that the unit is thoroughly dry, especially
around the mains power inlet, before reconnecting the power
cord and switching the unit on.
XXSEALER USER MANUAL - 1.0