7.3 Performing an xxplate Sealing Operation
A typical operating sequence to seal an
xxplate
is:
1) Switch the unit on and set the sealing time to 1.5 seconds and the
heater plate set-point temperature to 170ºC (using the method
described above). These are the recommended setting for the
xxplate
film.
2) If necessary, press the
HEATER
button to turn the heater on.
3) Allow sufficient time for the unit to reach temperature (the
Heater
LED will stop flashing and remain constantly on).
4) Fit the
BJS xxplate Adapter Block
into the
PCR 96 Well Skirted
Plate
Adapter Block
(both provided with the
xxsealer
) and locate
in the
Adapter Block Carrier
5) Load the
xxplate
onto the
xxplate Adapter Block
locating it onto
the sprung pins. It should sit flat on the adapter block.
And add the sealing film on top, the correct way up.
WARNING
Care should be taken not to touch
the surface of the heater plate whilst
loading the sample plate.