6 Unit Installation
Before installing the
xxsealer
unit, please check that the delivery is
complete and that the unit and any accessory parts are intact and free
from any signs of transportation damage. Also ensure that all external and
internal packaging has been removed from the unit before installation.
Please retain all packaging for future transportation and
storage of the unit and its accessories.
The
xxsealer
unit should be installed in a location which meets the
following requirements:
•
Safe and suitable operating environment (see Section 5)
•
Solid, stable, level working surface
•
At least 10cm clearance around the unit to adjacent objects and
walls
•
Earthed AC mains power connection (see Section 4)
WARNING
Please also observe and abide by the
Unit Installation and
Operating Environment
safety precautions and preconditions
listed in Section 5.
It is important the check that the Mains Voltage Selector switch on the rear
of the unit has been set to the correct position for the desired operating
voltage, and the correct value of fuse has been fitted (see Table 5).
Table 5: Mains Voltage Selection
Operating Voltage
110 to 120 VAC
220 to 230 VAC
Voltage Selection
See Section 3 for
switch location (
)
on unit rear panel.
115
230
Fuse Rating
3.15AH
1.6AH
XXSEALER USER MANUAL - 1.0