8 Maintenance and Servicing
Although the
xxsealer
unit does not require any scheduled maintenance
or servicing, the operator should regularly clean and inspect the unit for
any detects, as described in Section 8.3 below.
DANGER
WARNING
Please observe and comply with all of the
Unit
Maintenance and Serviceability
precautions listed in
Section 8.
Never removed the unit casework. There are no user or
operator serviceable parts inside the unit.
Always switch off and unplug the unit before performing
any cleaning or disinfecting tasks.
For technical and service related enquiries, please contact your distributor
or BJS Bio at the address given on Page 2 of this manual.
8.1 Replacing the Unit Fuse
The unit fuse should only be replaced by a suitably qualified technician.
DANGER
The unit fuse will only blow as a result of an internal unit fault
or if the voltage selector switch has been incorrectly set (see
Section 5). This fuse should only be changed after the unit has
been thoroughly inspected, and must be replaced with the
exact type specified in Section 4.
Thoroughly inspect the unit for any signs of damage, loose components
or liquid spillage or ingress. If in doubt, please contact BJS Bio on the
number given on Page 2 of this manual.
DANGER
The fuse holder is removed by disconnecting the mains cord
and then using a small flat bladed screwdriver to carefully pull
open the fuse holder (see Section 5) and remove the old fuse.
After replacing the fuse with the corrected rated one for the
operating voltage being used (see Table 5), push the fuse
holder firmly back into the inlet module.
The unit must be electrically safety tested for excess leakage
current before being repowered from the mains supply.