9
Special case single division scale 30000/10 g:
The device expects the input of the weighing range (steps 22, 25, 27) in
dimension [kg] or [lb]. The device expects the input of the digital incre-
ment (steps 23, 26, 28) in dimension [g] or [mlb].
If scales have a weighing range in dimension [g] but not [kg], select the
digital increment by
factor 1000 higher
.
Example:
Desired setting is 30000 g with scale interval 10 g.
Input: Step 22 =
30000
, step 23 =
10000
, step 29 =
0
.
9.3
Scale calibration menu
Slide switch to calibrated. No calibration possi-
ble.
The calibration process must be carried out with particular care.
You must first remove the verification seal to invalidate the verification
of the scale.
Menu call-up only via restart, see page 27.
Save data in EEPROM:
Press
key.
Fig. 26: Display: "AbGL_St"
Key
Display presentation
Process description
Call up service parameters and confirm with
.
Select calibration start and confirm with
.
Calibration point 0 preload.
Apply preload.
Display of measured value for calibration point 0.
First calibration point at 15 kg weighing range.
Example: 3 kg