BT-700 Operator Manual 18
P/N : 700-ENG-OPM-EUR-R02
Bistos Co.,Ltd.
2018,03
4.4
BT-700 Control Knob
In each mode, the control knob decrease and increase
the
pulse beat audio
volume
and push the knob, you can
enter the menu
.
In the other mode, the Control Knob is the method of adjusting parameters and
navigating through the menu system. If the knob is rotated while in a menu, the
cursor moves throughout the items within the menu. This process is used to
select a menu item for modification. The knob is then pressed to select this item
for editing.
Once a menu item has been selected for editing, the knob is rotated to scan
through the available choices for this parameter. Pressing the knob stores the
new value temporarily.
4.5
Menu Tree
The Menu Tree of BT-700 is such as table 4.4
Level 1
Level 2
Level 3
Default setting
System Setup
Target
Adult
Language
English
Alarms
SpO2 High/Low
limits (%)
Adult
H : 100 L : 85
Neonate
H : 95 L : 80
PR High/Low
limits (BPM)
Adult
H : 170 L : 40
Neonate
H : 190 L : 90
Alarm Temporary
Silence
30 s – 60s – 90s –
120s off
(default 120s)
Alarms Volume
3
Clock
Time Set
Time Format
Date Set
20yy/mm/dd
Patient Data
ID
00000
Name
AAAAAAAAAA
Trend Data
Data Interval
1 Min
List Type
Chart
Data List
Delete All Data
No
Table. 4.4 Trend Mode display
CAUTION
: Pressing the knob when “Memory All Delete”, saved data is
deleted.